Atc Absolute Temperature Calibration, 3-Point Calibration - Julabo F25-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F25-MC:
Table of Contents

Advertisement

Menu functions
9.2.

ATC Absolute Temperature Calibration, 3-point calibration

Examples:
1-point calibration
°C
T 1
T
3-point calibration
°C
T 1
T
T
1 = Original curve
T
1.
2.
40
ATC serves to compensate a temperature difference that
might occur between circulator and a defined measuring point
in the bath tank because of physical properties.
Principle:
For ATC calibration, in steady state the bath temperature
at the location of the temperature sensor (T
determined at the respective adjusted working
temperature. This value is then set on the circulator in the
menu >ATCalibration< under menu item >Calwert X<.
This can be a 1-point, 2-point or 3-point calibration.
°C
M = Temperature measuring instrument with
temperature sensor (TF)
B = Bath tank with T = circulator
T
= Temperature on measuring point
M
°C
T
= Temperature on circulator
T
The menu ATC - Absolute Temperature Calibration
is structured in 2 levels.
ATC status
Calibration type: 1-, 2-, 3-point calibration
2 values per calibration point
ttx = Temperature indication on circulator
Indicates the value of the calibration point entered last.
Ctx = Calibration temperature x
Input of the new value for the calibration curve.
) is
M