List Of Commands - Julabo F25-MC Operating Manual

Refrigerated and heating circulators
Hide thumbs Also See for F25-MC:
Table of Contents

Advertisement

Remote control

12.3. List of commands

out commands: Setting parameters or temperature values.
Command
version
status
out_mode_01
out_mode_01
out_mode_01
out_mode_05
out_mode_05
out_sp_00
out_sp_01
out_sp_02
out_sp_03
out_sp_04
out_sp_07
out_par_06
out_par_07
out_par_08
52
The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature >t 1< to 55,5 °C:
Command to ask for the working temperature >t 1<:
Response from the circulator:
Parameter
Response of circulator
None
Number of software version (V X.xx)
none
Status message, error message (see page 54 )
0
Use working temperature >t 1<
1
Use working temperature >t 2<
2
Use working temperature >t 3<
0
Stop the unit = R –OFF-.
1
Start the unit.
xxx.xx
Set working temperature. „t 1"
xxx.xx
Set working temperature. „t 2"
xxx.xx
Set working temperature. „t 3"
xxx.xx
Set high temperature warning limit „t High"
xxx.xx
Set low temperature warning limit
x
Set the pump pressure stage. (1 ... 4)
xxx
Xp control parameter of the internal controller.
xxx
Tn control parameter of the internal controller.
xxx
Tv control parameter of the internal controller.
out_sp_00 ⇔ 55.5↵
in_sp_00↵
55.5↵
„t Low"