8.4.

Change-over of the warning function to shutdown function

Factory setting: Li 0
t High
High temperature limit
t Low
Low temperature limit
For the two menu items > t High < and > t Low < choose
between a warning message being signalled or a complete
shutdown of the main functional elements such as heater
and circulating pump being effected. (see page 35)
• Setting >Li / 0<
An audible warning (interval tone) sounds and a
meassage appears on the MULTI-DISPLAY (LED).
t High
or
• Setting >Li / 1<
A complete shutdown of heater and circulating pump is
effected.
An audible alarm (continuous tone) sounds and a
message appears on the MULTI-DISPLAY (LED).
t High
or
1.
Press the key
is displayed.
2.
Press enter
the MULTI-DISPLAY (LED) (digit blinks).
3.
Select the parameter with the keys
4.
Press enter
Press
to update the display immediately.
t Low
t Low
.
until the menu item
to indicate the adjusted parameter on
to store the selected parameter.
MC
(0 / 1).
37