Preparations
5.4.
Temperature application to external systems
Caution:
If the circulator is operated without external system, close the pump
connector (24a) with the cap nut.
RS 232
15
A
M
P
15
A
M
P
24a
5.4.1. Tubing
Recommended tubing:
Order No. Length
8 930 008 1 m
8 930 010 1 m
8 930 108 1 m
8 930 110 1 m
8 930 410 1 m
8 930 209 0.5 m
8 930 210 1.0 m
8 930 211 1.5 m
8 930 214 3.0 m
8 930 220 0.5 m
8 930 221 1.0 m
8 930 222 1.5 m
8 930 223 3.0 m
*) Adapter for metal tubing M10x1 on M16x1
26
Securely attach all tubing to prevent slipping.
The circulator is used for temperature application to
external, closed systems (loop circuit) with simultaneous
temperature application in the circulator bath.
Connecting the external system
• Unscrew the collar nuts from the pump connector
(24a).
• Slide the tubing onto the pump connectors for feed
and return flow (24a, 24b).
24b
®
CR
tubing
®
CR
tubing
Viton tubing
Viton tubing
Insulation for tubing
Metal tubing, triple insulated, M16x1* -100 °C to +350 °C
Metal tubing, insulated, M16x1 *
8 mm inner dia.
10 mm inner dia.
8 mm inner dia.
10 mm inner dia.
8 mm or
10 mm inner dia.
Temperature range
-20 °C to 120 °C
-20 °C to 120 °C
-50 °C to 200 °C
-50 °C to 200 °C
-50 °C to 100 °C
-50 °C to +200 °C
Order No. 8 970 444