Menu functions
optimum setting
°C
Inappropriate settings may produce the following heat-up curves:
Xp too low
°C
Xp too high or Tv too high
°C
9.7.
Reset
Factory setting: no
46
Optimization instructions for the PID control parameters:
t
t
t
Use this to reset all values to factory setting.
1.
Press the
displayed.
2.
Press enter
3.
Press the keys
4.
Press enter
After the reset is carried out, the parameter is
automatically set to >no<.
The heat-up curve reveals inappropriate control
settings.
°C
Tv/Tn too high or Xp too high
°C
button until the menu item
to indicate the parameter (no).
to select the parameter (YES).
to confirm the reset.
Tv/Tn too low
t
t
is