8.2.

High and low temperature warning functions

Factory settings:
High temperature limit
t High
205 °C
Low temperature limit
t Low
-55 °C
If for a sensitive temperature application task adherence to a
working temperature value (setpoint) is to be supervised,
then set high and low temperature warning values.
In the example below, the >Setpoint< of 85 °C is surrounded
by the values > t High < 87 °C and > t Low < 83 °C. The
electronics immediately registers when the actual
temperature attains a temperature out of the limits and it
follows a reaction according to what is set in the menu item >
Li < (Warning or Alarm).
(see chapter 8.4. Change-over of the warning function to
shutdown function).
1.
Press the key
or
2.
Press enter
MULTI-DISPLAY (LED) (digit blinks).
3.
Set value:
Use the cursor keys
display until the numeral you wish to change blinks.
Use the increase/decrease arrows
selected numeral (-, 0, 1, 2, 3, ... 9).
4.
Press enter
Press
to update the display immediately
until the menu item
is displayed.
to indicate the adjusted value on the
to move left or right on the
to store the selected value.
MC
to change the
35