3.
4.
5.
4.2
Wire the power supply using an M12-
terminated cable (optional)
Use this procedure if you are using an M12-terminated cable to wire the power supply.
Installation Manual
Figure 4-1:
Location of power supply wiring terminals and equipment ground
A. Power supply wiring terminals (+ and -)
B. Equipment ground
Connect the power supply wires:
For DC power: connect to terminals + and –.
For AC power: connect to terminals L/L1 (line) and N/L2 (neutral).
Tighten the two screws holding the power connector in place.
Ground the power supply using the equipment ground, also under the Power
warning flap.
Power supply wiring