Wiring the channels
Figure 3-3:
A
A. Maximum resistance (Ω)
B. External supply voltage (V)
3.1.3
Wire the Frequency Output
Wire the Frequency Output in explosion-proof, nonincendive, or nonhazardous
installations.
Prerequisites
Important
Meter installation and wiring should be performed only by suitably-trained personnel using the
appropriate government and corporate safety standards.
Procedure
Wire to the appropriate output terminal and pins.
30
Externally-powered mA Output: maximum loop resistance
1100
1000
900
800
700
600
500
400
300
200
100
0
0
7.5
15.0
22.5
B
Micro Motion
®
Model 5700 Ethernet transmitters
30.0