3.1.6
Wire the I/O channel using an M12-terminated cable
(optional)
Use this procedure if you are using an M12-terminated cable to wire the I/O channel.
Prerequisites
Obtain an A-coded M12-terminated cable.
Procedure
1.
2.
Installation Manual
Attach the M12-terminated cable to the configuration I/O connector on the Model
5700 transmitter. See
Figure
Attach the other cable end using the pinouts described in
Figure 3-12:
M12-terminated cables to the Configuration I/O
Table 3-1:
M12 configuration I/O pinouts
Pin identifica-
tion
Wire color
Pin 1
Not used
Pin 2
White
Pin 3
Not used
Pin 4
Black
3-12.
Outputs
Not used
Positive terminal
Not used
Neutral terminal
Wiring the channels
Table
3-1.
35