B0048991 / Rev. 1.3
9.8
Faults concerning supply voltage
No.
Cause
60
Operating voltage was below 20 VDC.
62
Sensor voltage too low.
9.9
Faults Sensor Switch Box
No.
Cause
65
Error in the Sensor Switch Box.
9.10
Faults concerning PLC parallel inputs
No.
Cause
70
Operation of PLC inputs is incorrect.
71
The parallel function is not possible, because
another parallel function is active (e.g. measurement
start and in addition Teach-In 1-sheet).
72
Selected program number >255.
73
Parallel function is blocked in the system
configuration (ext. adjust: no).
76
The parallel function is not possible, because the
system was configured to "internal control".
77
The parallel function is not possible, because the
menu is open.
9.11
Other faults
No.
Cause
> 80
An error has occurred which cannot be
unequivocally identified to one of the above error
codes.
Pos: 65 /Komponenten allgemein/Leerseite @ 0\mod_1165223317234_501.doc @ 2343 @ @ 1
Pos: 66 /Kapitelübersicht/10. Wartung @ 0\mod_1121422080015_501.doc @ 514 @ 1 @ 1
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
Double Sheet Detector R1000 series UDK20
Manual
Fault messages, causes and remedies
Remedy
Check power supply.
Check external power supply
(insufficient power ?).
Contact Roland Electronic Service Department.
Remedy
If a Sensor Switch Box is connected, check box and
cables.
Otherwise read the notes in chapter 9.1.
Remedy
Check timing of PLC inputs.
See timing diagrams.
Select the new program number.
If required, change configuration.
In case of "internal control" it is only possible to select the
measuring program. If required, change configuration to
"external control".
Close the menu.
Remedy
Contact Roland Electronic Service Department.
105