Agilent Technologies N5242A Service Manual page 83

2-port and 4-port pna microwave network analyzers (10 mhz–26.5 ghz)
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5242A
Table 4-1
Power Supply Measurement Points
Test Point
Supply Name
TP517
Analog Ground
TP509
+15 V analog
TP511
+9 V analog
TP515
+3.3 V analog
TP513
+5.2 V analog
TP501
+15 V standby
TP504
+32 V analog
TP512
-15 V analog
If All Supply Voltages are Present
If all of the supplies have measured within tolerances, and the instrument still is not
functioning properly, refer to
If Any Supply Voltage Is Missing
Disconnect the line-power cord before removing any assembly.
WARNING
Procedures described in this document may be performed with
power supplied to the product while protective covers are removed.
Energy available at many points may, if contacted, result in personal
injury or loss of life.
You must sequentially remove all of the assemblies, taking care to disconnect the line
power cord before each removal, and then measure the supply voltages after each removal.
If the missing supply voltages return to a "power on" condition after removal of an
assembly, suspect that assembly as being defective.
Remove the network analyzer assemblies in the order specified in the following steps (refer
to
Chapter 7
for removal instructions).
1. Unplug the A19 test set motherboard ribbon cable from the A19 test set motherboard
(refer to
"Removing and Replacing the A19 Test Set Motherboard" on page
1. Unplug the A19 test set motherboard to A20 IF multiplexer board ribbon cable from
the A20 IF multiplexer board (refer to
Multiplexer Board" on page
2. Unplug the front panel interface cable from the A1 front panel interface board (refer to
"Removing the A1 Front Panel Interface Board and Keypad Assembly" on page
3. Remove the A12 SPAM board (refer to
on page
7-14).
4. Remove the A8 source board (refer to
page
7-14).
Service Guide N5242-90001
Expected
Test Point
Level (Vdc)
0 V
TP516
+15.0 ±0.1
TP514
+9.0 ±0.1
TP500
+3.3 ±0.1
TP502
+5.2 ±0.1
TP503
+15.0 ±0.1
TP507
+32.0 ±0.1
TP505
-15.0 ±0.1
TP519
"Front Panel Troubleshooting" on page
"Removing and Replacing the A20 IF
7-28).
"Removing and Replacing the A4–A13 Boards"
"Removing and Replacing the A4–A13 Boards" on
Troubleshooting
Power Up Troubleshooting
Expected
Supply Name
Level (Vdc)
-5.2 V analog
-5.2 ±0.1
-7 V analog
-7.0 ±0.1
Digital Ground
+5.1 V standby
+5.1 ±0.1
+12 V digital
+12.0 ±0.1
+3.35 V digital
+3.35 ±0.1
+5.1 V digital
+5.1 ±0.1
Digital Ground
4-13.
7-26).
0 V
0 V
7-10).
4-9