Agilent Technologies N5242A Service Manual page 112

2-port and 4-port pna microwave network analyzers (10 mhz–26.5 ghz)
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
WARNING
Procedures described in this document are performed with power
supplied to the product while protective covers are removed. Energy
available at many points may, if contacted, result in personal injury.
4. With the covers off, plug in the analyzer and turn on the power.
Do not operate the analyzer with the outer cover removed for more than
CAUTION
30 minutes, as this could cause the analyzer to overheat which could result in
costly damage.
Checking the A23 and A24 Mixer Brick Receiver Outputs
Set the network analyzer for an S11 measurement with a CW frequency of 1 GHz.
1. Locate the following flexible cables at the receiver channel IF outputs of the mixer
brick(s).
• 2-port models
— Channel A; W51 of A23
— Channel R1; W52 of A23
— Channel R2; W53 of A23
— Channel B; W54 of A23
• 4-port models
— Channel A; W51 of A23
— Channel R1; W52 of A23
— Channel R2; W53 of A23
— Channel B; W54 of A23
— Channel C; W55 of A24
— Channel R3; W56 of A24
— Channel R4; W57 of A24
— Channel D; W58 of A24
2. Disconnect the flexible cable at the suspect channel.
3. Connect the spectrum analyzer to the suspect channel connector.
4. The measured signal on the spectrum analyzer should be at 7.61 MHz.
• If the measured signal is present, continue testing at
Multiplexer Board" on page
5. If the measured signal is missing on the R1 channel, continue testing at
A33 Reference Channel Switch" on page
If the measured signal is missing on any channel (other than the R1 channel noted
above) and the analyzer does not have receiver attenuators, replace the A23 or A24
mixer brick, whichever is appropriate. Refer to
4-38
4-39.
4-39.
"Removing and Replacing the A23 and
PNA Series Microwave Network Analyzers
"Checking the A20 IF
"Checking the
Service Guide N5242-90001
N5242A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents