Install Backup Files onto the New Hard Disk Drive
The files that were previously saved onto a floppy disk or USB memory drive must now be
installed onto the new HDD. The network analyzer must be powered up and operating.
1. Insert the USB flash drive into a USB port.
2. Click on START. (It may be necessary to exit the Network Analyzer application.)
3. Point to Programs, Accessories, and then open Windows Explorer.
4. Navigate to C:\Program Files\Agilent\Network Analyzer.
5. Copy each of the backup files from the USB flash drive to the HDD.
6. Navigate to D:\CalFiles\.
7. Copy the backup files prefixed with mxcalfile_ from the USB flash drive to the HDD.
8. After all files have been copied, remove the USB flash drive from the USB port.
9. Exit Windows Explorer.
Post-Repair Procedures
1. Adhere the new Certificate of Authenticity (license) label to the outer cover in the
location specified in
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Service Guide N5242-90001
Figure 7-1 on page
7-7.
Table 7-2 on page
Repair and Replacement Procedures
Removing and Replacing the A51 Hard Disk Drive (HDD)
7-64.