Agilent Technologies N5242A Service Manual page 249

2-port and 4-port pna microwave network analyzers (10 mhz–26.5 ghz)
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5242A
4. The display cable (item
attach it to the LCD display hold down bracket (item
5. The LCD display can be removed by disconnecting the LCD cable from the inverter
board and then removing the four screws (item
hold down bracket (item
separating the LCD display from the hold down bracket.
Figure 7-5
A3 Display Assembly and Touch Screen Removal-2
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Service Guide N5242-90001
Removing and Replacing the A1–A3 and Other Front Panel Subassemblies
) can be removed by removing the three screws (item
). Note the location of the four spacers (item
Table 7-2 on page
Repair and Replacement Procedures
).
) that attach it to the LCD display
7-64.
) that
) before
7-13