Troubleshooting
Measurement System Troubleshooting
Checking the A5 and A8 Source Outputs
1. Refer to the block diagram at the end of this chapter and to
Options" on page
6-16. Locate the flexible cables W3 and W9 at the A5 source 1 board or
cables W5 and W7 at the A8 source 2 board.
2. Disconnect the cables, one at time, that are connected to cables W3 and W9 or cables
W5 and W7, from the source board to be checked.
3. Connect the spectrum analyzer to the open connector at the end of the cable that
connects to the source board to be checked.
4. Set the network analyzer for a 2 GHz CW frequency and observe the spectrum analyzer
measurement.
• The spectrum analyzer should measure a signal at 2 GHz (plus the 7.61 MHz offset).
5. If the signal is not present and the signals are present from
Source Synthesizer Outputs,"
and Replacing the A4–A13 Boards" on page
6. If the signal is present, reconnect all cables, and then continue with
Signal Separation Group" on page
4-32
replace the appropriate source board. Refer to
7-14.
4-33.
PNA Series Microwave Network Analyzers
"Top Cables, All Cables—All
"Checking the A4 and A13
"Removing
"Checking the
Service Guide N5242-90001