6.2

Programming the desired values

6.2.1 Setting the time

Fig. 15

6.2.2 Setting the automatic defrost time (cryochamber)

Fig. 16

6.2.3 Selecting the cryochamber temperature

Fig. 17
Leica CM1860 /CM1860UV – Cryostat
Use the function button labeled with a clock symbol to set the time of day.
To do so, set the current time using the
Pushing the
or
button for more than 1 sec. increases or decreases the
time continuously (auto-repeat function).
The automatic defrost cycle takes place once every 24 hours.
Briefly touch the
or
which has currently been set. The LED's between the indication of hours
and minutes will flash.
To change the beginning of the defrost cycle in 15-minute increments, press
the
or
button. The defrost time is 12 minutes.
The temperature of the cryochamber is set and indicated on the panel marked
with the cryostat symbol.
The actual temperature is the standard indication. Briefly touching the
or the
button displays the target temperature.
You can now set the desired value using these buttons. Pressing the
button for more than 1 sec. increases or decreases the chamber tem-
perature continuously.
The actual value will be indicated 5 seconds after finishing the programming.
6.
Operating the Instrument
and
button to display the beginning of the defrost cycle
keys.
or
31