Dispose of the UVC lamp separately!
Leica CM1860/CM1860 UV – Cryostat
3
5
8
9.
Cleaning, Disinfection, Maintenance
Removal of the lamp
The UVC lamp (3) is installed in front
of the protection screen for chamber
illumination (4).
• Hold the lamp with both hands and
carefully pull it out of the clips (5)
with a slight forward movement.
• Detach the metal ring (7) on the
holder (6) in the direction of the ar-
row (8) and carefully pull the lamp
Fig. 60
out of the holder with your right
hand
(see Fig. 60 +
Installation of the new lamp
• Carefully slide the metal ring (7) over
the lamp from the left
• Push the lamp into the holder on the
left side until it engages.
• Screw the metal ring onto the
holder, then hold the lamp with both
Fig. 61
hands and carefully push it into the
clips (5).
• Replace the sliding window.
• Connect the instrument to the power
supply again and switch it on.
Fig. 62
61).
(see Fig.
If any metallic mercury is
released, handle it carefully
and dispose of it properly!
62).
67