8.1

Error messages in the display

Error
Cause
20
Calibration error; possibly defective controller board.
21
Clock battery on the controller board empty.
23
Cryochamber temperature out of display range of
35 °C to -55 °C.
25
Temperature sensor for cryochamber temperature defective. Call service.
27
Temperature sensor for defrost limiter defective.
28
Both light barriers in the handwheel active.
29
Service interval warning after 15960 hours.
30
Service interval expired after 17610 hours.
If errors occur, they are output as error codes in the real-time display during normal operation
with the format EO:XX.
• In normal operation, the error displayed can be acknowledged by pressing any key once. Error
codes 21-28 are self-acknowledging if their cause has been eliminated.
• After 15,960 hours, "HELP" is output in the real-time display to indicate the need for service mainte-
nance. The display alternates between "HELP" and the real-time display. The note is acknowledged
by pressing a key on the keyboard. The "HELP" display reappears for a longer period.
• After 17,610 hours, "HELP" is output in the real-time display to indicate the need for service mainte-
nance. The display alternates between "HELP" and the real-time display. The note is acknowledged
by pressing a key on the keyboard. The "HELP" display reappears for a shorter period.
Leica CM1860/CM1860 UV – Cryostat
Error messages are displayed on the clock panel as follows EO: XX.
The following error messages might occur during operation:
Fig. 52
8.

Troubleshooting

Remedy
Switch the instrument off for
10 seconds and then back on.
If the error is displayed again:
Call service.
Call service.
Resolve cause.
Call service.
Call service.
Call service.
Call service.