Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 User Manual / Installation Instructions 10 716 079 – Version 01 Living up to Life...
Page 2
Thank you for choosing our products. We hope that you will enjoy the quality and performance of your Leica Microsystems surgical microscope. Leica Microsystems (Schweiz) AG Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 3
Controls Assembly Preparation before the operation General instructions Safety notes Care and maintenance What to do if...? Technical data Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Signs and labels Settings in the Leica MDRS for use with Leica FL800 Focussing and calibration of the NIR-data overlay Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Model: Serial No.: é Request for action; here, you are requested to take action. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Leica M720 OH5 FL800 Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 8
3 Leica OH5 illumination unit 4 Hinged arm monitor 5 MDRS 19“ touchscreen monitor (optional) 6 MDRS system unit (optional) Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 9
LED lights up in... Red = recording is on Brief flashing „Red“ = single-image recording 13 Leica DI C700 (optional) Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
45 S-video or BNC output, e.g. for OR monitor* 46 Keyboard (optional) *Only two of three outputs (37/43/45) can be used at the same time. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
(plug) in-between the enclosed Y-CAN cable. é Plug in the handswitch on the control unit at the internal CAN bus interface. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é Set the Leica M720 to minimum magnification. é Bring the monitor image into focus on the Leica Dual Video Adapter. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Use either only composite or only S-video signals for your installation. S-video signals feature a better quality than composite signals. If you are using S-video signals, connect the Y-cable of the FL800 NIR Sony camera with the BNC to S-video adapter to the FL connector board.
These settings can later be edited at any time via the User Settings menu. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 17
é Click on the „Save“ button to save your settings under the user name entered at the desired location. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
NIR-fine focus of the Dual Video Adapter at the chosen magnification. Fix the alignment with the enclosed adjustment screws. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 19
(Service setting only for this calibration). Refer to the user manual Leica M720 OH5. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 20
é The calibration was successful if both fluorescence objects overlay the injected fluorescence information injected into the Leica DI C700. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
NIR dual video adapter right = display side left For setting up the optional Leica DI C700 refer to the user manual Leica M720 OH5. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
The status LED on the swinging arm of the Leica OH5 is illuminated in yellow if the FL800 mode is active, and in white if switched back to white-light mode. Furthermore the status LED for recording is illuminated in red during the NIR recording.
Page 23
The photo function is available only during the live image (not in playback mode)! The playback of the recording can ensue in normal speed, slow motion or image per image. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é The display of the FL800 video signal is carried out automatically upon activation of the FL800 mode or when starting the playback of an FL800 loop. é The display of the FL800 video signal is carried out in front of a black background. The second beam path can optionally be open and closed.
45 S-video or BNC output, e.g. for OR monitor* 46 Keyboard (optional) *Only two of three outputs (37/43/45) can be used at the same time. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
(plug) in-between the enclosed Y-CAN cable. é Plug in the handswitch on the control unit at the internal CAN bus interface. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é Set the Leica M525 to minimum magnification é Bring the monitor image into focus on the Leica Dual Video Adapter Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Use either only composite or only S-video signals for your installation. S-video signals feature a better quality than composite signals. If you are using S-video signals, connect the Y-cable of the FL800 NIR Sony camera with the BNC to S-video adapter to the FL connector board.
These settings can later be edited at any time via the User Settings menu. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 35
In the service menu, you can adjust a value for the maxi- mum duration of the FL800 mode from 10 to 180 seconds. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
NIR-fine focus of the Dual Video Adapter at the chosen magnification. Fix the alignment with the enclosed adjustment screws. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 37
(Service setting only for this calibration). Refer to the user manual Leica M525 OH4. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 38
é The calibration was successful if both fluorescence objects overlay the injected fluorescence information injected into the Leica DI C500. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é Switch on the illumination. See also QRG Leica FL800 for Leica M525 OH4. Sterility é Apply the sterile drape. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é A connected monitor will display the NIR image of the ICG test pattern. Check the alignment of the FL800 NIR Sony camera (see page 29). Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
The status LED on the swinging arm of the Leica OH4 is illuminated in yellow if the FL800 mode is active, and in white if switched back to white-light mode. Furthermore the status LED for recording is illuminated in red during the NIR recording.
Page 42
The photo function is available only during the live image (not in playback mode)! The playback of the recording can ensue in normal speed, slow motion or or image per image. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é The display of the FL800 video signal is carried out automatically upon activation of the FL800 mode or when starting the playback of an FL800 loop. é The display of the FL800 video signal is carried out in front of a black background. The second beam path can optionally be open and closed.
Page 44
Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Leica M525 F50 FL800 Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 46
9 Hand grip 10 Base 11 HDMD or MDRS documentation system (optional) 12 Camera Control Unit (optional) 13 Control unit Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 47
Red = recording is on Brief flashing „Red“ = single-image recording 13 Optional Leica DI C500 14 White-light camera Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
45 S-video or BNC output, e.g. for OR monitor* 46 Keyboard (optional) *Only two of three outputs (37/43/45) can be used at the same time. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
(plug) in-between the enclosed Y-CAN cable. é Plug in the handswitch on the control unit at the internal CAN bus interface. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é Set the Leica M525 to minimum magnification é Bring the monitor image into focus on the Leica Dual Video Adapter Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Head External Monitor HD-Camera Control Unit HDMD Recording Unit HDMD Monitor F50 Control Unit HDMDTouch Pad F50 MDC Board Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 53
An overview of all components connected to your system has been provided in your delivery documenta- tion by your Leica representative. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
These settings can later be edited at any time via the User Settings menu. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 55
In the service menu, you can adjust a value for the maxi- mum duration of the FL800 mode from 10 to 180 seconds. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
NIR-fine focus of the Dual Video Adapter at the chosen magnification. Fix the alignment with the enclosed adjustment screws. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 57
(Service setting only for this calibration). Refer to the user manual Leica M525 F50. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 58
é The calibration was successful if both fluorescence objects overlay the injected fluorescence information injected into the Leica DI C500. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é Switch on the illumination. See also QRG Leica FL800 for Leica M525 F50. Sterility é Apply the sterile drape. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é A connected monitor will display the NIR image of the ICG test pattern. Check the alignment of the FL800 NIR Sony camera (see page 49). Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
The status LED on the swinging arm of the Leica F50 is illuminated in yellow if the FL800 mode is active, and in white if switched back to white-light mode. Furthermore the status LED for recording is illuminated in red during the NIR recording.
Page 62
The photo function is available only during the live image (not in playback mode)! The playback of the recording can ensue in normal speed, slow motion or or image per image. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
é The display of the FL800 video signal is carried out automatically upon activation of the FL800 mode or when starting the playback of an FL800 loop. é The display of the FL800 video signal is carried out in front of a black background. The second beam path can optionally be open and closed.
é Connect the VGA Cable 10715599 to the Connector Board. Gui- de the cable through the cabel holder along the cable bundle (1). Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 65
é Click the Y-Cable FL800 and the VGA Cable into the cable guide the red markings (1). bar 1074363 (1). Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 66
1074363 flush with the red marking and pull on the woven sleeve 10743363 (1). é Add the woven sleeve 10743310. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 67
é Guide the Y-Cable FL800 in parallel with the Panasonic Camera Cable and fix with cable ties (1). é Final look. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 68
Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
The standard preventive measures and safety regulations per- taining to electromagnetic and other radiation have to be observed. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
• Other accessories, provided that these have been expressly approved by Leica as being technically safe in this context. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
The Leica M720 OH5 FL800 / M525 OH4 FL800 surgical microscope is intended for use in the electromagnetic environment specified below. The customer or the user of the Leica M720 OH5 FL800 / M525 OH4 FL800 surgical microscope should assure that it is used in such an environment.
Page 73
The Leica M525 F50/C50 FL800 surgical microscope is intended for use in the electromagnetic environment specified below. The customer or the user of the Leica M525 F50/C50 FL800 surgical microscope should assure that it is used in such an environment.
Page 74
The Leica M720 OH5 FL800 / M525 OH4 FL800 surgical microscope is intended for use in the electromagnetic environment specified below. The customer or the user of the Leica M720 OH5 FL800 / M525 OH4 FL800 surgical microscope should assure that it is sed in such an environment.
Page 75
The Leica M525 F50/C50 FL800 surgical microscope is intended for use in the electromagnetic environment specified below. The cus- tomer or the user of the Leica M525 F50/C50 FL800 surgical microscope should assure that it is used in such an environment.
Page 76
The Leica M720 OH5 FL800 / M525 OH4 FL800 surgical microscope is intended for use in the electromagnetic environment specified below. The customer or the user of the Leica M720 OH5 FL800 / M525 OH4 FL800 surgical microscope should assure that it is used in such an environment.
Page 77
The Leica M525 F50/C50 FL800 surgical microscope is intended for use in the electromagnetic environment specified below. The cus- tomer or the user of the Leica M525 F50/C50 FL800 surgical microscope should assure that it is used in such an environment.
Page 78
If different accessories or other cables than specified here or allowed by the manufacturer of the Leica M720 OH5 FL800/ M525 OH4 FL800 surgical microscope are used, it may lead to increased electromagnetic radiation or to a decrease in immunity to distur- bances.
Page 79
Warning message: The Leica M525 F50/C50 FL800 surgical microscope may not be used while positioned directly next to other instruments. If it is necessary to operate them in the vicinity of other instruments, the microscopes should be monitored to ensure that they function properly in this arrangement.
• oxygen splitting compounds. Follow the disinfectant manufacturer’s instructions. It is recommended to conclude a service contract with Leica Service. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
• Dry off products, except for optical components, with a dispos- products for their effectiveness and possible disadvantages. able cloth/paper cloth. Dry optical surfaces with a micro-fibre cloth. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 83
Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Cause 1: You already recorded 12 loops é C reate a new recording folder – For additional information, see the MDRS4 User Manual. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
–40 °C to +70 °C –40 °F to +158 °F 10% to 100% rel. humidity 500 mbar to 1060 mbar atmospheric pressure Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
The Leica M720 OH5 FL800 / M525 OH4 FL800 / M525 F50 FL800 sur- gical microscope may be used only in closed rooms and on solid level floor. Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 87
Technical data Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 88
Technical data Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 89
Technical data Leica FL800 for Leica M720 OH5, Leica M525 OH4 and Leica M525 F50 / Ref. 10 716 079 / Version 01...
Page 90
The statement by Ernst Leitz in 1907, “with the user, for the user,” describes the fruitful collaboration where we rank with the market leaders. with end users and driving force of innovation at Leica Microsystems. We have developed five brand values to live up to this tradition: Pioneering, High-end Quality, Team Spirit, Dedication to •...