Daily Use of the Instrument
Blade holder CE
4
23
11
Knife holder CN
39
Disinfection
Place all of the removed parts into a disinfect-
ing solution for the time recommended by the
manufacturer. Dry each part thoroughly before
reassembly and insertion into the cryostat.
Occasionally oil the moving parts such as the
axes (37) and (38) and the slot (39) with a drop of
cryostat oil (Type 407).
Leica CM1860/CM1860 UV – Cryostat
10
2
12
36
Fig. 49
37
38
Fig. 50
1. To release the segment arc (2) from the base,
use a 4 mm Allen key to loosen the screw
over the clearance angle adjustment (23) and
remove the segment arc (2) from the blade/
knife holder base.
2. Fold the anti-roll guide system (4) to the left
while holding the lever (11).
3. Open the clamping lever (10,
it counterclockwise and pull it out.
4. Afterwards, the pressure plate (36) can be
removed for cleaning (with alcohol).
5. Open the clamping lever of the blade holder
(12,
Fig.
49) by turning it counterclockwise and
pull it out. The blade holder can now be moved
sideways and can be taken out of the segment
arc.
If several blade/knife holders are
cleaned at the same time, the parts
must NOT be mixed up! Failure to ad-
here to this may result in sectioning
problems!
Also observe the warning message on
Page 13!
Fig.
49) by turning
49