Measurement Speed Control Commands - Keysight V3500A User Manual

Handheld rf power meter
Hide thumbs Also See for V3500A:
Table of Contents

Advertisement

N O T E
Command Name
Description
Syntax
Response
Comments

Measurement speed control commands

Command Name
Description
Syntax
Response
Comments
V3500A User's Guide
Sending a value using this command results in taking averages of 2
measurements (where x = <Value>).
For example:
• Sending SETAVG<1><LF> causes the average factor to take 2
measurements
• Sending SETAVG<2><LF> causes the average factor to take 2
measurements
• Sending SETAVG<3><LF> causes the average factor to take 2
measurements
When <Value> = 0, (sending SETAVG 0), the number of measurements
is set to 1, and is equivalent to averaging disabled.
Get averaging factor
This command reads and returns the averaging factor, which
is equal to log
(number of measurements per average).
2
AVG?<LF>
<Value>
A value of 0 is equivalent to having averaging turned off.
Set normal speed mode
This command sets the V3500A to normal speed mode.
NMODE<LF>
OK
Normal speed mode is the default mode of operation.
The V3500A accuracy specifications apply in normal mode.
Measurement speed is level- dependent.
Driver Installation and Commands
3
x
1
or 2
2
or 4
3
or 8
57

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents