Figure 2-12 Commands That Control Measurement Time - Keysight U2722A User Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

Operation and Features
2
However, due to the latency of approximately 15 ms, the total time taken for the
measurement to complete and return the value would be approximately 5.115 s.
Hence, it is recommended that the time-out settings for communicating with the
U2722A/U2723A is set to more than 5.115 s, else time-out error would likely
occur.
Controlling measurement samples
You can vary the number of data points in a measurement sample and the time
between samples. Refer to
Figure
2-12.
SENS:SWE:TINT <time>, (@1|2 |3)
SENS:SWE:POIN <# of points>, (@1|2 |3)
Figure 2-12
Commands that control measurement time
When the instrument is turned on and at *RST, the output voltage or current
sampling period is calculated using the SENSe:SWEep:TINTerval <integer>,
(@1|2|3) command and the measurement buffer size is set to a value ranging
from 1 up to 4096 data points. Thus, the expected time taken to fill up the data
points set in the data buffer is governed by the following formula,
×
Expected time, t = points
time interval
You can vary the data sampling rate with these commands.
SENS:SWE:TINT <sample_period>, (@1|2|3)
SENS:SWE:POIN <points>, (@1|2|3)
Keysight U2722A/U2723A User's Guide
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

U2723a

Table of Contents