Chapter 3 Key Functions
Average
(2-25)
Back
3-16
Sets the number of the sweeps to be averaged. The number
of the sweeps can be adjusted from 2 to 25 using the
Up/Down arrow key, or entered directly with the keypad
number keys. Select OFF, or enter 1 to turn off averaging.
Press ENTER to accept the selected value.
This feature uses a measurement routine that averages sam-
ples over a user-defined number of successive sweeps. The
averaging is done on a point-by-point basis. Each displayed
point reflects the current measurement associated with that
point averaged with the measurements associated with that
point from previous sweeps. This feature can be used to min-
imize the impact of transients on the displayed signal and
noise.
Min
Displays and holds the minimum responses of the input sig-
Hold
nal over multiple sweeps. This feature is useful for detecting
small, discrete, stable signals in the presence of noise.
Returns to the previous menu.