Transducers; Temperature Sensor - Siemens SITRANS LUT400 Operating Instructions Manual

Ultrasonic controllers
Hide thumbs Also See for SITRANS LUT400:
Table of Contents

Advertisement

Transducers

Warning: Hazardous voltage present on transducer terminals during
operation.
Notes:
Do not connect coaxial cable directly to SITRANS LUT400 due to high voltage
transmitted on shield of coaxial cable
Do not connect the LUT400 shield and white transducer wires together; wire to
separate terminals
Disregard older transducer manuals that recommend these practices

Temperature sensor

The speed of sound changes as temperature changes. To ensure accurate level
measurement, the SITRANS LUT400 compensates via an external temperature input. All
Siemens EchoMax transducers have an internal temperature sensor for this purpose,
and for the fastest temperature response, Siemens also offers a dedicated temperature
sensor, the TS-3.
If the following conditions apply, a separate TS-3 temperature sensor will ensure
optimum accuracy:
the transducer is exposed to direct sunlight (or other radiant heat source)
the temperature of the atmosphere between the transducer face and monitored
surface differs from the temperature of the transducer
faster response to temperature changes is required.
To achieve the best performance of temperature measurement in a typical open channel
flow application, the temperature sensor should be shielded from direct sunlight and
mounted half way between the ultrasonic transducer face and the maximum head
achievable in the application. Care should be taken to not obstruct the direct sound path
of the ultrasonic transducer.
Use a TS-3 Temperature Sensor only. Leave terminals open (unused) if TS-3
Note:
is not deployed.
A5E33329501
TS-3
1
2
SITRANS LUT400 (HART) – OPERATING INSTRUCTIONS
TRANSDUCER
3
4
5
Page 23