S E C T I O N 8 • O p e r a t i o n

Check Disposables Mode

The Check Disposables mode of the Fluid Warmer indicates a missing
or improperly installed Disposable Set.
Mode characteristics
• The Check Disposables yellow LED (a) on the Fluid Warmer's
Display Panel is illuminated.
• An audible attention signal beeps.
• Reservoir solution circulation is stopped; fluid warming stops.
• Pressure Chambers continue to operate.
To Clear this mode
Install a disposable or check the disposable installation as follows:
• Check the position of the disposable in the #1 Block.
• Make sure the heat exchanger is seated in the heat exchanger guide.
• Press down firmly on the #2 Block.
• Check the position of the Gas Vent/Filter Assembly in the #4
Interlock.

Add Recirculating Solution Mode

The Add Recirculating Solution mode of the Fluid Warmer indicates
that the solution level in the recirculating solution reservoir is below its
minimum level.
Mode characteristics
• The yellow Add Recirculating Solution LED (
Warmer's Display Panel, is illuminated
• An audible warning signal beeps.
• Solution circulation stops, fluid warming stops.
• Pressure Chambers continue to operate.
• Fluid flow to the patient continues.
To Clear this mode
Add recirculating solution to the Fluid Warmer's reservoir.
32
), on the Fluid
b
| Operator's Manual