S E C T I O N 4 • I m p o r t a n t S a f e t y I n f o r m a t i o n
WARNINGS for the Air Detector/Clamp [continued]
Do not turn OFF the Fluid Warmer when the Air Detector alarm
is active. If the Fluid Warmer is powered OFF in an active alarm
state, the Air Detector/Clamp will open and the Air Detector
will become disabled. This could allow any air within the patient
line to be delivered to the patient resulting in serious injury or
death.
The functional test for the Air Detector/Clamp accessory
must be performed before each use. If any visual indicator
does not illuminate or the audible signal does not sound,
do not use the Fluid Warmer. Remove the device from service
immediately. Fully functional visual and audible alarm systems
are essential for the safe use of the Air Detector/Clamp.
Malfunction, failure, or damage to the device may occur if
these cautions are not followed:
Do not use the Fluid Warmer if equipment or Disposable
Set malfunction is evident.
When loading fluid bags into Pressure Chambers, choose a
hanging hook that allows the bag port to hang freely in the
indented slot at the bottom of the chamber door. If bag ports
are positioned above this slot, diminished flow could occur.
Federal law (U.S.A.) restricts this device to sale by or on the
order of a physician.
8

CAUTIONS

| Operator's Manual