9.3 Replace the Gas Vent/Filter Assembly
Use one of the following Gas Vent/Filters Assemblies:
• F-10 Gas Vent / Filter Assembly for DI-50, D/DI-60HL, D/DI-70,
and D/DI-100 Disposable Sets
• F-30 Gas Vent / Filter Assembly for D/DI-300 Disposable Sets
• F-50 Gas Vent / Filter Assembly for DI-55, D/DI-65HL, D/DI-75,
D/DI-150, and D/DI-350 Disposable Sets
Close all clamps above and below the Gas Vent/Filter Assembly on
1
D/DI-series Disposable Set and on new Gas Vent/Filter Assembly.
Turn the Fluid Warmer OFF.
2
Steps 3 and 4 are different for the H-1200 and H-1000 Fluid
Note:
Warmers. Continue with the appropriate steps 3 and 4 that apply for
your specific Fluid Warmer.
Remove the
3
used
the Disposable Set:
Remove the Gas Vent/Filter Assembly from the #4 Block.
a
For DI-55, D/DI-65HL, D/DI-75, D/DI-150, and D/DI-350
disconnect the clip from the bottom of the Heat Exchanger.
Open the #3 Clamp Slot door and remove the patient line from
b
the Clamp Slot.
Install the
Gas Vent/Filter Assembly:
4
new
The steps to install the Gas Vent/Filter Assembly differ
Note:
depending on the Disposable Set. Continue with the appropriate
steps that apply for your Disposable Set.
Use the following steps for DI-55, D/DI-65HL, D/DI-75,
D/DI-150, and D/DI-350.
Press the
a
new
press the clip to the bottom of the Heat Exchanger.
Insert patient line in the Clamp Slot.
b
Align the patient line in the Clamp Slot without kinking.
c
Close the Clamp Slot door.
d
Go to step 5 to continue.
e
Use the following steps for DI-50, D/DI-60HL, D/DI-70,
D/DI-100, and D/DI-300.
Insert patient line in the Clamp Slot.
a
Close the Clamp Slot door.
b
Align the patient line in the Clamp Slot without kinking.
c
Press the Gas Vent/Filter Assembly into the #4 Block.
d
Go to step 5 to continue.
e
48
Gas Vent/Filter Assembly while still connected to
Gas Vent/Filter Assembly into the #4 Block and
| Operator's Manual
F-50
F-10/30