F-50
F-10/30
S E C T I O N 9 • O p e r a t i n g I n s t r u c t i o n s
Remove the
3
used
while still connected to the Disposable Set.
For DI-55, D/DI-65HL, D/DI-75, D/DI-150, and D/DI-350
disconnect the clip from the bottom of the Heat Exchanger.
4 Press the
Gas Vent/Filter Assembly into the #3 Block.
new
For DI-55, D/DI-65HL, D/DI-75, D/DI-150, and D/DI-350
press the clip to the bottom of the Heat Exchanger.
5 Using aseptic technique:
Disconnect the upper Luer fitting on the
a
Assembly.
Remove the upper Luer end cap on the
b
Assembly.
Connect the Disposable Set to the
c
inlet.
6 With the lower clamp on the
open all clamps above Gas Vent/Filter Assembly.
The Gas Vent/Filter Assembly will self prime.
Turn the power ON.
7
Remove the end cap on the lower patient line of the
8
Vent/Filter Assembly. Slowly release the lower clamp and allow the
lower patient line to fill completely.
Close the clamp after the line is full.
9
Holding the
10
used
the Luer lock on the lower patient line. Connect the Luer-lock
fittings on the
used
Connect the Luer lock on the lower patient line of the Gas
11
Vent/Filter Assembly to the Patient Line.
12 Open clamps below the Gas Vent/Filter Assembly and resume
infusion.

9.4 Activated Alarms

Refer to Section 8, Operation for information on identifying Alarm states,
conditions that activate them, and methods for clearing Alarm states.
A. "Air Detected" Alarm mode
Detection of air in the Gas Vent/Filter Assembly results in the
following:
• The patient line is clamped off
• The red Clamped LED warning signal illuminates
• The audible warning signal beeps
Gas Vent/Filter Assembly from the #3 Block
used
new
Gas Vent/Filter Assembly
new
Gas Vent/Filter Assembly closed,
new
Gas Vent/Filter Assembly horizontally, disconnect
Gas Vent/Filter Assembly together and discard.