Slow Flow Rate Troubleshooting Guide

Problem
Check the following:
Pressure Chambers
Verify the Fluid Warmer is ON.
not activated
Verify the Pressure Chamber levers are fully in the plus (+),
pressurized position.
Verify pneumatic tubing is fully seated into fittings on the
Pressure Chambers and Fluid Warmer.
Blood develops
Stored blood develops particulate. If blood is older
particulate
than 5-7 days, consider using Level 1
Clogged filter
Change Gas Vent/Filter Assembly if filter becomes
clogged.
Fluid bag not fully
Make sure the membrane of the fluid bag port is fully
spiked or bag port
pierced by the bag spike and that the neck of the bag port
twisted
is not twisted.
®
Non-Level 1
bag
40 to 80 micron filters used between the bag port and
port filters
the spikes of the Disposable Set may restrict flow. Consider
using Level 1
Clamps partly
Verify all clamps are fully open.
engaged
Clamps left in the
Leaving clamps fully clamped for long periods will cause
clamped position
the tubing to become deformed. Do not leave clamps
for long periods
closed for extended periods of time.
Kinked tubing
Verify that no tubing kinks are present.
Air trapped on filter
Remove Gas Vent/Filter Assembly and tap against the
screen of Gas Vent/
cabinet of the Fluid Warmer to dislodge air bubbles,
Filter Assembly
allowing them to vent. If this fails to correct problem, replace
with new Gas Vent/ Filter Assembly.
Non-high flow
Use only extension lines with an inner diameter of
extension lines
0.13" (3.3mm) or larger, such as Level 1
extension sets.
Stopcock
Use only Level 1
Small gauge needle
Use large bore needles or catheters to maximize flow
or catheter
rates.
S E C T I O N 1 0 • T r o u b l e s h o o t i n g
®
PF-1 pre-filters (340 micron).
®
SC-3 (9 french inner diameter).
| Operator's Manual
®
PF-1 pre-filters.
®
X-36 or Y-30