Fluid Warmer Display Panel

The Fluid Warmer Display Panel provides continuous information
about the operation of the Fluid Warmer. A liquid crystal display (LCD)
indicates recirculating solution temperature. Just below the LCD, four
light-emitting diodes (LEDs) indicate operation modes for the device.
For identification purposes, the diodes are shown illuminated.
1
2
3
4
5
a
Reservoir Level Display
The Reservoir Level Display has a clear window for viewing the amount
of recirculating solution present in the reservoir. Check the reservoir to
ensure the solution level is near the maximum level indicator (
recirculating solution level is too low, the Add Recirculating Solution
LED on the Display Panel illuminates and an audible attention signal
beeps.
S E C T I O N 7 • C o n t r o l s a n d D i s p l a y s
Recirculating Solution Temperature
The temperature of the recirculating solution is displayed in the
LCD panel. The temperature is displayed in degrees celsius.
This is NOT the temperature of fluid delivered to the
Note:
patient—the display reflects the temperature of the recirculating
solution.
Automatic Operation LED
The green LED indicator illuminates when the power is ON and
the Disposable Set has been properly installed. When lit, this
indicates the Fluid Warmer is operating.
Check Disposables LED
The yellow LED indicator illuminates and an audible attention
signal beeps when the Disposable Set is not properly installed.
See the Interlocks description in this section for directions on
clearing the Check Disposables alarm.
Add Recirculating Solution LED
The yellow LED indicator illuminates and an audible attention
signal beeps when reservoir is low. Additional recirculating
solution must be added to the reservoir. Maximum capacity for
the reservoir is 1.4 liters of recirculating solution.
Over Temperature LED
The red LED indicator illuminates and an audible warning signal
beeps when the recirculating solution is over the acceptable
temperature for safe use.