S E C T I O N 9 • O p e r a t i n g I n s t r u c t i o n s
Blood and blood products could contain pathogenic organisms.
Failure to follow Institutional policy and procedures for
biomedical-hazardous materials could lead to exposure to
harmful pathogens which could result in user death or serious
injury.
Step 3—Make Patient Connection
Make patient connection and begin infusion.
Step 4—Replace the Gas Vent/Filter Assembly
Replace the Gas Vent /Filter Assembly every 3 hours, when the
filter becomes clogged, or if air is venting slowly. Refer to Section
9.3, Replace the Gas Vent/Filter Assembly.
Step 5—Change the Fluid Bag
Move the lever on the Pressure Chamber over to the
a
minus (–) unpressurized position. This will release the pressure in
the Pressure Chamber and deflate the bladder.
Close pinch clamp under empty bag.
Open door and remove the fluid bag from the Pressure Chamber.
b
Remove the spike from the used fluid bag.
c
Remove any air from the new fluid bag and spike the fluid bag.
d
Hang the new fluid bag in the Pressure Chamber. Close and latch
e
the door.
Move the lever on the Pressure Chamber over to the
f
plus (+) pressurized position to pressurize the chamber.
Open pinch clamp.
g