Over Temperature Alarm Mode

The Over Temperature Alarm mode is entered when the temperature of
the recirculating solution reservoir is at or above
A
ctivation of the Over Temperature warning signal indicates that
warming has stopped and immediate operator intervention is
required. Failure to clear the over temperature condition or to
take the device out of service may result in patient death or
serious injury.
Mode characteristics
• The Over-Temperature LED warning light illuminates (
• An audible warning signal beeps.
• Solution circulation is stopped; fluid warming stops.
• Pressure Chambers continue to operate.
• Fluid flow to the patient continues.
To Clear Over Temperature Alarm mode
Do the following:
• If connected to a patient, close all clamps.
• Turn OFF the Fluid Warmer to clear the alarm.
• Turn the power back ON.
Power ON Test for Air Detector/Clamp
The
Power ON Test
the Fluid Warmer is turned ON. This test activates the Air
Detector/Clamp's visual and audible indicators.
Mode characteristics
• Audible alarm indicator beeps.
• All LED indicators on the Air Detector/ Clamp Control Panel
illuminate:
1 Automatic Operation LED indicator - Green
2 Check Tubing LED indicator - Yellow
3 Clamped LED indicator - Red
At the end of the Power ON Test the Air Detector/Clamp enters
Automatic Operation mode. This is the default mode for the Air
Detector/Clamp.
S E C T I O N 8 • O p e r a t i o n
WARNING!
for the Air Detector/Clamp is activated when the
| Operator's Manual
43.9°C
).