S E C T I O N 1 2 • M a i n t e n a n c e
4 Rinse a separate soft cloth or sponge in room temperature
running potable water. Squeeze out excess water so that the
applicator is not dripping. Wipe all of the aforementioned
surfaces. Repeat rinsing the cloth or sponge several times with
fresh running water during this process to insure all visible
residue is removed.
5 Dry the item with a hand towel or soft cloth.
6 Visually inspect the Fluid Warmer and its components to insure
that they have been thoroughly cleaned. Repeat cleaning
procedure if necessary.
7 After thoroughly cleaning the Fluid Warmer, perform
disinfection if required.
8 If it is hospital policy to perform disinfection as part of
reprocessing, then follow your institution's guidelines for
disinfecting of the surfaces of non-critical medical devices. The
list below includes low-level disinfectants that are commonly used
in the medical community and high-level disinfectants that are
claimed by the manufacturer. The effectiveness of these listed
disinfectants should be validated using the hospital procedures.
The following disinfectant agents can be used without causing
damage to the enclosure:
Low Level Disinfectants:
• fantastik
High Level Disinfectants:
• 1.59% Phenol (e.g., Sporicidin
• 3.4% Glutaraldehyde (e.g., CIDEX
• 10% Bleach solution
• 1% Ammonia solution
• Surface disinfectants compatible with plastic or metal materials
9 Rinsing of the disinfectant residue should be done using a
soft cloth or sponge as the applicator.

General Inspection

• Check the condition of the Fluid Warmer with a visual inspection.
Remove from service any unit that shows physical damage.
• If the Disposable Set does not install easily, lubricate the O-Rings
as directed in the following section.