NOTE:
Chamber pressure must be released before the pressure chamber door
can be opened.
To Clean
Please refer to the maintenance section of this manual for complete cleaning
and disinfecting procedures for the System 1025 pressure infusers.
Disposables: Set Up and Performance
| WARNING!
Refer to D-series disposable packaging for complete set up and
priming
instructions, indications and contraindications warnings, and cautions, and
additional product information.
Setting Up Disposable Sets
SIMS Level 1 D-series disposable sets have been designed to be highly
effective and easy to use in combination with the System 1000 fluid warming
hardware.
By following the simple 1-2-3 blocks found on the System 1000 fluid
warmer, properly trained users can confidently have the system safely set-up
and operating within seconds.
Following the numbers found on the System 1000:
1. Push the bottom end of the heat exchanger into the bottom heat exchanger
#1 socket. Be sure the heat exchanger is fully seated.
! WARNING !:
t
Do not bend heat exchanger tube. Bending the heat exchanger
tube may damage the inner aluminum tube.
If the tube is bent,
serious patient injury or death could result.
2. Slide the top +2 socket up. Snap the heat exchanger into the heat exchanger
guide.
Be sure it is fully seated.
Slide the top heat exchanger socket down over
the heat exchanger tube.
If the top socket is not pressed down correctly, a red indicator next to the top
socket will illuminate.
When the top socket is properly seated the red indicator
will go out.
3. Snap the Filter with Gas Vent into the #3 holder on the front of the cabinet.
4. Press the green power switch on the front of the cabinet to turn ON the unit
12