S E C T I O N 7 • C o n t r o l s a n d D i s p l a y s
Air Detector/Clamp Control Panel and Alarms
The Air Detector/Clamp Control Panel has three LED indicators that
display the operational state of the Air Detector/Clamp. Refer to figure
on right.
1
Automatic Operation LED
The green Automatic Operation LED illuminates when the
following conditions are present: The Fluid Warmer power is
ON, a Disposable Set is properly installed in the Fluid Warmer
and primed, the patient line from the Gas Vent/Filter Assembly
is correctly placed in the #3 Clamp Slot, and the Clamp Slot
door is closed.
2
Check Tubing LED
The yellow Check Tubing LED illuminates and an audible
attention signal beeps when the patient line tubing from the Gas
Vent/Filter Assembly is not correctly placed in the #3 Clamp
slot, and when the Clamp Slot door is not closed correctly.
3
Clamped LED
The red Clamped LED illuminates and an audible warning signal
beeps when air is detected in the Gas Vent/Filter Assembly. The
patient line is automatically clamped.
24
| Operator's Manual