S E C T I O N
8

Operation

The Level 1
®
H-1200 Fast Flow Fluid Warmer (Fluid Warmer) performs three
primary functions; Fluid Warming, Air Detection/Clamping, and Pressurized
Fluid Delivery. Functions are monitored and controlled by four
interfaces/control panels located on the Fluid Warmer.
The four interfaces are:
• Fluid Warmer Power and Alarm Test Panel
• Fluid Warmer Display Panel
• Air Detector/Clamp Alarm and Control Panel
• Pressure Chamber Control Panel
In the table on the facing page, operation of the device is represented in terms
of the four interfaces that control specific device functions. The numbers call-
out individual Modes of Operation activated or indicated on the interface.
Functions
Fluid Warming
Air Detection/Clamping
Pressurized Fluid Delivery
Operational modes
OFF Mode
ON/Automatic Operation for Fluid Warmer
Alarm Test Mode
Over Temperature Test Mode
Check Disposables Mode
Add Recirculating Solution Mode
Over Temperature Alarm Mode
Power ON Test for Air Detector/Clamp
Automatic Operation Air Detector/Clamp
Check Tubing Mode
Air Detected/Clamped Mode
Pressurized Mode
Unpressurized Mode
The modes of operation are individually defined in the following section. This
includes a description of each mode, activation and/or monitoring of the
mode, mode characteristics, and clearing of the mode state.
S E C T I O N 8 • O p e r a t i o n
| Operator's Manual
29