Alarm Test Mode
The Alarm Test mode is used to test the visual and audible indicators of
the Level 1
®
H-1200 Fast Flow Fluid Warmer. This mode is entered by
pressing and holding the Alarm Test button (
Control Panel.
Mode characteristics
• All visual indicators on the Fluid Warmer's Display Panel (e)
illuminate.
• The Fluid Warmer's audible alarm beeps.
• When the Alarm Test button is released, the Over Temperature
LED and an audible alarm remains active.
• To clear the Over Temperature alarm, turn the Fluid Warmer OFF
and then back ON.
Over Temperature Test Mode
The Over Temperature Test mode is used to test the operation of the
Fluid Warmer's Over Temperature Circuitry. This mode is entered by
pressing and holding the Over Temperature Test button (
Warmer Control Panel with the Fluid Warmer at operating temperature
(41°C).
Mode characteristics
• The red Over Temperature LED (
illuminates.
• An audible warning signal beeps.
To Clear this mode
• Turn Fluid Warmer OFF.
• Turn Fluid Warmer back ON.
Temperature Display
The Temperature Display functions when the Fluid Warmer is powered
ON. Temperature is displayed in degrees Celsius.