Specifications - level 1 H-1200 115V Operator's Manual

Fast flow fluid warmer
Table of Contents

Advertisement

S E C T I O N
1 5

Specifications

System Specifications
Physical
Height, Overall
Width, Overall
Depth, Overall
Weight Assembled; Dry
Recirculating Solution Capacity
Air Source Pressure
Environmental
Operation
Transportation
Storage
Thermal
Temperature Set Point
Over Temperature Set Point
Electrical
Protection Against Electrical Shock
Mode of Operation
Type of Current
Ingress Protection Rating
MAINS Power Input
MAINS Power Input
Auxiliary MAINS Outlet
Auxiliary MAINS Outlet
S E C T I O N 1 5 • S p e c i f i c a t i o n s
Measurement
67 inches
20 inches
20 inches
63 pounds
0.37 gallons
300 (294 ± 6) mm/Hg
Temperature
10°C to 40°C
-18°C to 60°C
5°C to 40°C
41.7°C ± 0.3°C
43.9°C ± 0.1°C
Type
Class I Equipment
Type BF Equipment
Continuous
Alternating
IPX1
115 VAC, 60 Hz, 12 Amps
230 VAC, 50 Hz, 6.3 Amps
115 VAC, 1.5 Amps
230 VAC, 0.75 Amps
| Operator's Manual
(1.7 m)
(51 cm)
(51 cm)
(28.5 kg)
(1.4 L)
Humidity [%]
10 to 95
5 to 95
5 to 95