Change Recirculating Solution With 0.3% Hydrogen; Peroxide/Distilled Water Solution; Change O-Rings; Clean Fan Filter - level 1 H-1200 115V Operator's Manual

Fast flow fluid warmer
Table of Contents

Advertisement

S E C T I O N 1 2 • M a i n t e n a n c e
9 Turn the Fluid Warmer ON. Let the solution circulate
for a 30-minute disinfection period.
10 Turn the Fluid Warmer OFF.
11 Empty the reservoir.
12 Remove the Disposable Set and discard according to established
hospital procedures
Change Recirculating Solution with a 0.3% Hydrogen

Peroxide/Distilled Water Solution

1 Place a container under the drain valve of the Fluid Warmer.
2 Drain the recirculating solution by turning the drain valve
clockwise 90 degrees.
3 When all the solution has drained from the reservoir, close the
drain valve.
4 Prepare a 0.3% hydrogen peroxide/distilled water solution. Mix
140 ml of 3% hydrogen peroxide solution and 1,260 ml of
distilled water.
5 Fill the reservoir with 0.3% hydrogen peroxide/distilled water
solution. The reservoir holds 1.4 liters.

Change O-Rings

Change the O-Rings in the #1 Block and #2 Block.
1 Remove each O-Ring from its socket by pulling it out
with a pair of needle-nose pliers or by prying it out with
a small screwdriver.
2 Lubricate the new O-Rings from the O-Ring Kit.
3 Press each O-Ring into its socket.

Clean Fan Filter

The Fan Filter (
) is located on the bottom of the Fluid Warmer.
a
The fan guard snaps in place.
1 Remove four screws and unsnap the fan guard from the bottom
of the unit.
2 Clean the filter with warm soapy water.
3 Replace the fan guard and filter.
62