S E C T I O N 9 • O p e r a t i n g I n s t r u c t i o n s
6
Prime Drip Chamber by squeezing drip chamber until one-half
to three-quarters full of fluid.
• DI-50, DI-55, D/DI-60HL, D/DI-65HL, D/DI-70, and
Note:
D/DI-75 series use a single drip chamber.
• D/DI-100, D/DI-150, D/DI-300, and D/DI-350 series
use a separate drip chamber for each spiked bag.
The clamp is below the drip chamber.
7
Open remaining clamps above the Heat Exchanger. Fluid flows into
the Gas Vent/Filter Assembly.
8
For DI-50, D/DI 60HL, D/DI-70, D/DI 100, and D/DI 300
vigorously tap the Gas Vent/Filter Assembly to dislodge air bubbles
from filter screen.
9
Press the green Power ON button, located on the Power and Alarm
Test Panel, to turn ON the Fluid Warmer.
• The Air Detector/Clamp runs a Power ON Test.
• All Air Detector/Clamp indicator LEDs illuminate.
• The audible warning beeps.
– If the above does not occur, remove the device from service.
• Upon completion of the Power On Test the Air Detector/Clamp
enters Operation mode with the Automatic Operation LED
illuminated.
• If the Disposable Set is incorrectly installed, the Fluid Warmer's
Check Disposables attention indicator illuminates and the
audible attention signal beeps. Check the installation of the
Disposable Set following the directions provided in Section 6
Controls and Displays , on Interlocks.
C—Prime the Patient Line
Remove the male Luer cap from the distal end of the Patient Line.
1
On Level 1
Note:
Administration Sets verify that no recirculating solution
comes out of the distal end of the Patient Line.
42
®
D/DI-60HL and D/DI-65HL Disposable
| Operator's Manual

Advertisement

Table of Contents
loading