TEMPERATURE circuitry. Press and hold this button to test this
alarm circuitry. When the switch is released the WATER OVER
TEMPERATURE alarm will continue to sound. This is normal
operation.
To clear the alarm condition, turn the unit OFF and then
back ON again.
Figure 3. Power and Alarm Test Panel
Figure 4. Display Panel
C
x
|
C
y
ON POWER OFF
@PERATING
|
O
BISPOSABLE
ALARM
TEST OVERTEMP
ABB WATER
DI @VERTEMP
1.4 Liters
Distilled Water Only
q
|
q
|
y
Display Panel
The display panel shown in Figure 4 contains visual information that continuously
indicates the status of the System 1000 during operation.
Five displays provide
information for safe, convenient use of the device.
AAA
1.
CIRCULATING
WATER
BATH
TEMPERATURE
Displays temperature of the circulating water in °C. Note that this is not the
temperatu! ire of the fluid delivered
to the natient,
but the temperature
of the
heat
SOUTCE.
2. OPERATING
This GREEN indicator illuminates when the power switch is ON and the heat exchanger
has been properly installed.