Bottles And Sizes-Standard; Suction Line - Teledyne 5800 Installation And Operation Manual

Refrigerated sampler isco
Hide thumbs Also See for 5800:
Table of Contents

Advertisement

Isco 5800 Refrigerated Sampler
Section 3 Programming
FIRST SAMPLE AT:
08:00 16-JUN-12
Times —
When entering times, use the 24-hour format,
sometimes known as "military time." In this format,
times after 12:00 noon continue to increment (1:00 pm
= 13:00, 2:00 pm = 14:00, etc.), until midnight or 00:00.
3.3.2 Bottles and
Sizes-Standard
SELECT OPTION: (<-->)
BOTTLES AND SIZES
NUMBER OF BOTTLES:
1
2
4
BOTTLE VOLUME IS
___ ml (300-65500)
IS A DISTRIBUTOR ARM
IN USE? YES NO
BOTTLE FULL DETECT?
YES NO
CUSTOM DISCHARGE
TUBE? YES NO
DISCHARGE TUBE
LENGTH: __._inches

3.3.3 Suction Line

SELECT OPTION: (<-->)
SUCTION LINE
3-4
About Times and Dates
Select this option to specify the installed bottle configuration
(section 2.4). The sampler uses this setting to determine
available bottle options and control the operation of the
distributor.
1. Select the BOTTLES AND SIZES option and press Enter.
2. Press the Previous or Next arrows to highlight the
installed bottle configuration. Press Enter to select it.
24
3. Press the Number buttons to enter the bottle volume in
milliliters. For multi-bottle samplers, enter the bottle
volume of one bottle, not the total volume of all bottles.
If you enter a non-standard bottle volume, the sampler will
ask "ARE YOU SURE?". Answer NO to re-enter a standard
volume or answer YES to confirm the non-standard
volume.
4. For the one-bottle, 10L configuration, answer YES. For the
one-bottle, 20L configuration, select NO.
a. If you selected NO distributor arm present, select YES
to make the sampler stop the program if the bottle
becomes full, NO to continue running the program.
5. If you are using a discharge tube length other than the
standard 17.5" and 20" (such as the 20L bottle
configuration), select YES.
6. Enter the custom tube length and press Enter.
Select this option to enter the suction line length (section 2.5).
The sampler uses this input along with the pump tables it
creates to determine the duration of the pump fill and purge
cycles and to calculate sample volumes.
To set the suction line length:
1. Select the SUCTION LINE option and press Enter.
ENTER TIME AND DATE:
HH:MM
DD-MON-YY
Dates —
When entering dates, use a day-month-year format.
Although the screens above display the date as a
three-character abbreviation and a two-digit numeric
value, months are always entered by typing the
numeric value (1=January, 2=February, and so on).