4.3.3 Sampling Reports

4.3.4 Remove Sample
Bottles
All times are reported as serial values, which are easily
converted to dates and times by spreadsheet programs. The
numbers to the left the decimal point are the number of days
since January 1, 1900. The numbers to the right are fractional
days which can be converted to hours, minutes, and seconds.
To retrieve a report of the sampler's program settings, type
SETTINGS<CR>.
The sampler records the sample and bottle number at the time of
each sample event. A sampling results report can be obtained by
typing RESULTS<CR>. The RESULTS report contains data
from the most recently run program only.
For a report containing both settings and results, type
REPORT<CR>.
For a report containing both SETTINGS and RESULTS from the
Previous Run, type PREVIOUS REPORT<CR>.
For a report containing RESULTS from the Previous Run, type
PREVIOUS RESULTS<CR>.
For a report containing SETTINGS from the Previous Run, type
PREVIOUS SETTINGS<CR>.
After the program is complete, the bottles can be removed and
prepared for the laboratory. Gain access to the bottles by
releasing the door latch and swinging the door open. The bottles
can then be removed from the rack or locating base. Push the
distributor arm out of the way if needed.
If you are using the 24-bottle rack, you can slide it out to easily
reach the bottles in the back. When sliding the rack out, observe
the rack guide pins (Figure 4-3). The back two guide pins run
along the top of the rack to prevent the rack from tipping
forward. Do not pull the rack out beyond these guide pins unless
you are securely holding the rack.
Removed bottles should be capped and labeled according to the
protocol established by your laboratory.