5800 Refrigerated
Sampler
Installation and Operation Guide
Part #69-5803-086 of Assembly #60-5804-051
Copyright © 2012. All rights reserved, Teledyne Isco
Revision H, August 2016

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 5800 and is the answer not in the manual?

Questions and answers

Summary of Contents for Teledyne 5800

  • Page 1 5800 Refrigerated Sampler Installation and Operation Guide Part #69-5803-086 of Assembly #60-5804-051 Copyright © 2012. All rights reserved, Teledyne Isco Revision H, August 2016...
  • Page 3 Teledyne Isco recommends that you read this manual completely before placing the equipment in service. Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction.
  • Page 5 Isco 5800 Refrigerated Sampler Isco 5800 Refrigerated Sampler Safety Information The Isco 5800 Refrigerated Sampler is a “definite purpose” device, intended for use only with compatible Isco equipment. Except as described in this manual, do not use this product with any other manufacturers’...
  • Page 6 Isco 5800 Refrigerated Sampler General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed in the following general...
  • Page 7 Isco 5800 Refrigerated Sampler This manual applies Hazard Severity Levels to the safety alerts, Hazard Severity Levels These three levels are described in the sample alerts below. CAUTION Cautions identify a potential hazard, which if not avoided, may result in minor or moderate injury. This category can also warn you of unsafe practices, or conditions that may cause property damage.
  • Page 8 Isco 5800 Refrigerated Sampler The equipment and this manual use symbols used to warn of Hazard Symbols hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual.
  • Page 9: Table Of Contents

    2.2 5800 Sampler Installation Overview ........
  • Page 10 Isco 5800 Refrigerated Sampler Table of Contents 3.3.10 Program Options ..........3-10 3.3.11 Language .
  • Page 11 A.1 5800 Refrigerated Sampler ........
  • Page 12 2-12 5800 Sampler to Isco flow meter cable ....... .
  • Page 13: Sampler Overview

    • Features, page 1-2 • Specifications, page 1-4 1.1 Sampler Overview The 5800 Refrigerated Sampler is an automated wastewater sampler that can collect liquid samples from a variety of sources including open channels, sewers, and storm pipe conduits. This indoor/outdoor use sampler is made of durable UV-resistant plastic that provides exceptional resistance to corrosion and weathering.
  • Page 14: Features

    Isco 5800 Refrigerated Sampler Section 1 Introduction 1.2 Features The 5800 Series sampler features are identified in Figure 1-1 and described in Table 1-1. Front View Left View Back View Figure 1-1 5800 Sampler features...
  • Page 15 Isco 5800 Refrigerated Sampler Section 1 Introduction Table 1-1 5800 Sampler Features Item Name Description (Fig. 1-1) Control Panel Cover Protects the control panel display and keypad. Can be locked with user supplied padlock. Control Panel Supports user control and programming of sampler operation.
  • Page 16: Specifications

    Isco 5800 Refrigerated Sampler Section 1 Introduction 1.3 Specifications Table 1-2 Specifications 132  73  84 cm 52.0  28.8  33.2 in Size (HWD): Weight (empty): 88.5 kg 195 lb Bottle Configurations: 13 configurations available: 24, 1-liter PP...
  • Page 17 Isco 5800 Refrigerated Sampler Section 1 Introduction Table 1-2 Specifications (Continued) Flow Meter Signal Inputs: 5 to 15 volt DC pulse or 25 millisecond isolated contact closure for Isco flow meters. 4-20 mA input for non-Isco flow meters. Digital Alarms: 4 programmable outputs;...
  • Page 18 Isco 5800 Refrigerated Sampler Section 1 Introduction...
  • Page 19: Section 2 Installation

    AC power source. Contact Teledyne Isco or its authorized representative for advice. Compare the contents of the boxes with the enclosed packing slip. If there are any shortages, contact Teledyne Isco immediately. 2.2 5800 Sampler The following sections provide general instructions for placing the sampler into operation.
  • Page 20 Isco 5800 Refrigerated Sampler Section 2 Installation WARNING If this product is used in a manner not specified in this manual, the protection provided by the equipment may be impaired. WARNING This product has not been approved for use in hazardous locations as defined by the National Electrical Code.
  • Page 21: Minimum Space Requirements

    Isco 5800 Refrigerated Sampler Section 2 Installation thereby reducing the possibility of sample cross-contamination. Refer to Table 1-2 for maximum suction line lengths and suction head heights. • Security – The sample compartment and pump and control panel have a padlock feature, but the location may need additional security to prevent tampering or vandalism.
  • Page 22 Isco 5800 Refrigerated Sampler Section 2 Installation Note To level the sampler, place a bubble level on the installed bottle rack or locating base, then adjust the feet. For one-bottle configurations, place the bubble level on the side rails. CAUTION Do not tip the sampler on its side or back.
  • Page 23: Connecting Power

    This delay allows time for the compressor heater to drive off any refrigerant dissolved in the compressor oil. 2.4.1 Backup Power The 5800 is capable of switching the sampler controller to backup power from an Isco lead-acid battery in the event of AC power loss. Note This provides backup power to the sampler controller only;...
  • Page 24: Installing A Bottle Kit

    5800 battery backup module 2.5 Installing a Bottle Kit The 5800 sampler can hold 1, 2, 4, and 24 bottles. Each of these bottle configurations are supplied as a kit (see Table 2-1), which is normally ordered with the sampler or when you desire to change the bottle configuration.
  • Page 25: Installing The Distributor Arm And Discharge Tube

    1-bottle Configuration. Includes one glass 2.5-gallon (10-liter) round bottle, two PTFE lined caps, locating base, and two discharge tubes. 68-5800-018 Single ProPak Configuration for 5800 and 6700 Series FR Refrigerated Sampler. Includes one composite ProPak holder with cap, 100 ProPak bags, two discharge tubes and an instruction sheet. 68-5800-011 1-bottle Configuration.
  • Page 26: Bottles

    Then, retest the tube at other arm positions. Note You can order replacement discharge tubes from Teledyne Isco. Order part number 60-9003-279 for a single tube. You can also order bulk tubing rolls from which you can cut to 20 inches (51 cm).
  • Page 27: Insert The Keyed End First So That It Faces The Back Of The Refrigerated Compartment

    Isco 5800 Refrigerated Sampler Section 2 Installation 1. Place the uncapped bottles in the rack (Figure 2-3). 2. If you are using 350 ml glass bottles, compress the retaining ring and place it in the center of the rack. Release the ring and the outward force of the ring will ®...
  • Page 28: The 24-Bottle Rack Is Held In Place By A Latch On The Left Side

    Isco 5800 Refrigerated Sampler Section 2 Installation Figure 2-4 The 24-bottle rack is held in place by a latch on the left side 2-10...
  • Page 29: 2.5 Gal) Bottles

    Isco 5800 Refrigerated Sampler Section 2 Installation 2.5.3 4 10L (2.5 Gal) Bottles The four bottle configuration uses a locating base to hold and align the bottles. To install the base and bottles: 1. Insert the four uncapped bottles into the numbered holes in the locating base 2.
  • Page 30: 4, 20L (5 Gal) Bottles

    Isco 5800 Refrigerated Sampler Section 2 Installation 2.5.5 4, 20L (5 Gal) Bottles 1. Place the four bottles in the locating base with the bottle mouths positioned above the holes in the base. 2. Slide the locating base into the bottom of the refrigerated compartment.
  • Page 31: One 20 Liter Bottle Configuration With Discharge Tube Through The Bottle Cap

    Isco 5800 Refrigerated Sampler Section 2 Installation you place the bottle in any other position liquid will miss the bottle. 20 liter bottle To install the 20 liter bottle: The distributor arm should be removed to prevent it from being an obstruction while you replace the bottle.
  • Page 32: Attaching The Suction Line

    The sampler uses a -inch ID suction line of Line lengths from 3 to 99 feet. Teledyne Isco offers vinyl or FEP-lined suction lines. The FEP-lined tubing has a polyethylene jacket to protect it from kinks and abrasions.
  • Page 33: Fep-Lined Suction Line

    Isco 5800 Refrigerated Sampler Section 2 Installation Figure 2-8 Attaching the suction line to the pump tubing Once the coupler is attached to the pump tube, removal is difficult, and may require cutting the tube. 2.6.2 FEP-lined Suction Line Inserting a tube coupler into FEP-lined suction line will damage the thin lining.
  • Page 34: Attaching A Strainer

    Isco 5800 Refrigerated Sampler Section 2 Installation 2.7 Attaching a Strainer To select the right strainer for your application, see Table 2-2 and Figure 2-10. To attach the strainer to the suction line, carefully screw the strainer’s threaded connector into the suction line. If attaching the strainer to an FEP-lined suction line, heat the end of the suction line before threading the strainer into the line.
  • Page 35: Alternative To Strainers

    Isco 5800 Refrigerated Sampler Section 2 Installation 2.7.1 Alternative to Strainers When sampling from high velocity streams with heavy suspended solids, some field investigations suggest that more representative samples are obtained without the strainer. Consider attaching a short piece of thin walled aluminum tubing to the end of the suction line;...
  • Page 36: Connecting The Sampler To External Devices

    Isco 5800 Refrigerated Sampler Section 2 Installation 2.9 Connecting the You can connect the sampler to an external instrument (Figure 2-11) for a variety of reasons. Typical reasons include: Sampler to External Devices • Receiving flow pulses from a flow meter device for variable time or variable volume sample collection.
  • Page 37: Other Device Connections

    All of these types of connections use the optional unterminated connect cable, P/N 68-5800-020 (Figure 2-13). This 3 m (10 ft) cable is unterminated at the device end to allow wired connections to compatible circuits. Refer to Table 2-3 for wire identification and connection details.
  • Page 38 Isco 5800 Refrigerated Sampler Section 2 Installation Figure 2-14 Quick disconnect box 60-2004-228 Table 2-3 Unterminated Connect Cable Sampler Pin Wire Color Signal Name Input/Output Parameters/Comments Black 12 VDC Power Output +14.5 to +12.0 VDC, 0 to1 ampere maximum. Power for external devices.
  • Page 39: Configuring And Programming The Sampler

    Isco 5800 Refrigerated Sampler Section 2 Installation 2.10 Configuring and To complete the installation, the sampler software should be configured and programmed. Refer to Section 3 for instructions. Programming the Sampler Configure the sampler to make sure that it “knows” what bottle configuration is installed, the length of the suction line, etc.
  • Page 40 Isco 5800 Refrigerated Sampler Section 2 Installation 2-22...
  • Page 41: Section 3 Programming

    Isco 5800 Refrigerated Sampler Section 3 Programming In this section… • Control Panel Description • Getting Started, page 3-2 • Configuring the Sampler, page 3-3 • Sampling Program Overview, page 3-11 • Programming Steps, page 3-15 • Programming Examples, page 3-24 3.1 Control Panel...
  • Page 42: Getting Started

    Isco 5800 Refrigerated Sampler Section 3 Programming Table 3-1 Control Button Functions Icon Name Description Power Places the sampler in the On or Standby modes. Note: In either modes, mains power is always con- nected to the refrigeration system. Numbers Types a number.
  • Page 43: Configuring The Sampler

    (section 4.3) will not contain meaningful data until you have run a sampling program. 3.3 Configuring the Before operating the 5800 Sampler, configure the sampler software. Doing so will set the time and date, and allow the Sampler sampler controller to correctly use the hardware and external connections.
  • Page 44: Bottles And Sizes-Standard

    Isco 5800 Refrigerated Sampler Section 3 Programming About Times and Dates FIRST SAMPLE AT: ENTER TIME AND DATE: 08:00 16-JUN-12 HH:MM DD-MON-YY Times — Dates — When entering times, use the 24-hour format, When entering dates, use a day-month-year format.
  • Page 45: Liquid Detector

    Isco 5800 Refrigerated Sampler Section 3 Programming 2. Press the Number buttons to enter the suction line length. SUCTION LINE LENGTH: ___ ft (3-99) 3. The sampler creates pump tables before returning to the PLEASE WAIT! SELECT OPTION <– –> screen.
  • Page 46: Flow Mode Sampling

    Isco 5800 Refrigerated Sampler Section 3 Programming 3.3.5 Flow Mode Sampling Select this option to indicate whether or not the sampler should take a sample when the Run Program button is pressed. The Sample At Start option is only applicable to Flow Paced programs.
  • Page 47: Refrigeration

    Isco 5800 Refrigerated Sampler Section 3 Programming 3.3.7 Refrigeration Select this option to set the targe t temperature of the refrigeration system. The sampler will operate the cooling or heating systems as needed to maintain this target temperature. 1. Select the REFRIGERATION option and press Enter.
  • Page 48 Isco 5800 Refrigerated Sampler Section 3 Programming • PUMPING FORWARD – This event occurs for the duration of the forward pumping step of a sampling event (between pre- and post-purge). • PUMPING REVERSE - This event occurs for the duration of the reverse pumping steps of a sampling event (pre- and post-purge).
  • Page 49: Tubing Life

    Check the condition of the pump tube regularly and if the tube shows signs of fatigue or wear, replace it immediately. A properly maintained sampler will provide years of reliable service that is expected of a Teledyne Isco Sampler.
  • Page 50: Program Options

    ____ them again. Should you forget the password, contact Teledyne Isco’s Customer Service department for assistance. 3.3.11 Language Select this option to change the display language and display units for length measurements.
  • Page 51: Run Diagnostics

    Isco 5800 Refrigerated Sampler Section 3 Programming 3.3.13 Run Diagnostics Select this option to enter the sampler diagnostics mode to test the sampler memory, display, keypad, pump, distributor, and various inputs and outputs. For more information about this service-related feature, refer to Section 5.2.
  • Page 52 Isco 5800 Refrigerated Sampler Section 3 Programming Single-bottle samplers can only distribute the sample to the composite bottle therefore skip the steps in this category. • Volume instructions define how much liquid is collected at each interval, and the total volume to collect.
  • Page 53: Time Paced Sampling Program Flow Chart

    Isco 5800 Refrigerated Sampler Section 3 Programming PROGRAM CONFIGURE VIEW LOG TIME PACED 3.5.2 FLOW PACED SAMPLE EVERY 3.5.3 __ HOURS, __ MINUTES One Bottle Multiple Bottles __ BOTTLES PER 3.5.5 SWITCH SETS EVERY SAMPLE EVENT (1-max) 3.5.7 __ HOURS, __ MINUTES SWITCH ON TIME 3.5.6...
  • Page 54: Flow Paced Sampling Program Flow Chart

    Isco 5800 Refrigerated Sampler Section 3 Programming PROGRAM CONFIGURE VIEW LOG TIME PACED 3.5.2 FLOW PACED SAMPLE EVERY 3.5.4 __ PULSES (1-9999) One Bottle Multiple Bottles __ BOTTLES PER 3.5.5 SWITCH BOTTLES EVERY SAMPLE EVENT (1-max) 3.5.7 __ HOURS, __ MINUTES SWITCH ON TIME 3.5.6...
  • Page 55: Programming Steps

    Isco 5800 Refrigerated Sampler Section 3 Programming 3.5 Programming Steps To begin programming from the Main Menu screen, use the Next or Previous arrow to select PROGRAM. Press the Enter button to display the first programming screen. Refer to Figures 3-3 or 3-4 and the following descriptions.
  • Page 56: Sample Every

    Isco 5800 Refrigerated Sampler Section 3 Programming Use the Next or Previous arrows to select the time or flow option, then press the Enter button. 3.5.3 Sample Every Time Paced Only. Use the Number buttons to enter the time __ Hours, __ Minutes interval in hours and minutes.
  • Page 57 Isco 5800 Refrigerated Sampler Section 3 Programming Table 3-2 Flow Pulse Intervals at Various Input Currents Input Seconds Between % of Full Scale Current (mA) Pulses Flow Rate 17.4 68.75 14.8 81.25 13.7 87.5 12.7 93.75 To calculate the number of flow pulses to enter in this sampler programming screen, follow the steps below.
  • Page 58: Switch On Time Or Number Of Samples

    Isco 5800 Refrigerated Sampler Section 3 Programming Example 1. Consider a flow meter programmed to output 20 mA at 1,000 liters per minute, the peak flow rate of the channel. If you are required to collect a sample every 10,000 liters, you would enter 50 flow pulses according the equation.
  • Page 59: Switch Bottles/Sets Every

    Isco 5800 Refrigerated Sampler Section 3 Programming 3.5.7 Switch Bottles/Sets Multiple Bottles Only. If you selected Switch on Time (section Every __ Hours, __ 3.5.6), use the Number buttons to enter the desired time interval Minutes for bottle or bottle set switches.
  • Page 60: Enter Sample Volume Use Defined Sample

    Isco 5800 Refrigerated Sampler Section 3 Programming If you select NO, the sampler will run until it completes the distribution sequence, at which time it stops the program and reports PROGRAM DONE. The sampler will wait in this state until the bottles have been emptied and the program is restarted.
  • Page 61: Ml For Every

    To measure the suction head height, refer to Figure 3-5. For most applications, Teledyne Isco recommends that you turn on the liquid detector (Section 3.3.4). The sampler will automatically calculate the suction head height, typically resulting in more accurate and repeatable sample volumes.
  • Page 62: Pause & Resume

    Isco 5800 Refrigerated Sampler Section 3 Programming Figure 3-5 Suction Head Height 3.5.19 Pause & Resume The Pause & Resume option allows you to pause the operation of the sampler and then resume operation at a later time and/or date. This option has to be enabled before the sampler is programmed.
  • Page 63: Set Start Time

    Isco 5800 Refrigerated Sampler Section 3 Programming Screen you will see when no pairs are set up First screen where you enter in the first pair of Pause & Resume times and days of the week Example of a Pause & Resume...
  • Page 64: Auto Re-Run

    Isco 5800 Refrigerated Sampler Section 3 Programming Note If you enter a “First Sample At” or Start Flow Count At” month less than the current month, the sampler interprets the start time as next year’s date. 3.5.21 Auto Re-Run Use this option when you want to sample on specific days of the week, like Monday, Wednesday, Friday.
  • Page 65 Isco 5800 Refrigerated Sampler Section 3 Programming • Flow-proportional Constant Time Variable Volume Program Example: Defining the Sample Volume If you have programmed the sampler to USE DEFINED deposit it into the bottle. To measure the delivered SAMPLE (section 3.5.11), follow the steps in this volume, you can replace the bottle with a example to define the volume using pump counts.
  • Page 66 Isco 5800 Refrigerated Sampler Section 3 Programming Example: Time Paced Sampling Program This example shows how to program the sampler to 7. The sampler displays the Run Continuously collect a 500 ml sample every hour for twenty-four screen. Select NO to stop the sampling program hours.
  • Page 67 Isco 5800 Refrigerated Sampler Section 3 Programming Time Paced Sampling Program, continued 13. The sampler displays the Select Days screen. The 15. The sampler displays the Auto Re-Run screen. day or days when the program is set to run will be Select NO and press Enter.
  • Page 68 Isco 5800 Refrigerated Sampler Section 3 Programming Example: Flow Paced Sampling Program This example shows how to program the sampler to 5. The sampler displays the Bottle Switch screen. Our collect a 100 ml sample every 10,000 gallons of liquid program requires a bottle switch every 24 hours.
  • Page 69 Isco 5800 Refrigerated Sampler Section 3 Programming Flow Paced Sampling Program, continued 12. The sampler displays the Start Flow Count screen. 14. The sampler displays the First Valid Day screen. Type “00” and press Enter, then type “00” and Enter the calendar day for the program to begin...
  • Page 70 Isco 5800 Refrigerated Sampler Section 3 Programming Example: Flow-proportional Constant Time Variable Volume Program This example shows how to program the sampler to Flow screen. Select YES and press Enter. collect flow-proportional samples volumes at fixed time VOLUMES DEPENDENT intervals. This program will collect a sample every 15 ON FLOW? minutes.
  • Page 71: Section 4 Operation

    Isco 5800 Refrigerated Sampler Section 4 Operation In this section… • Start a Sampler Program • Pause or Stop a Running Program, page 4-3 • Post Sampling Activities, page 4-3 • View the Log, page 4-3 • Retrieve the Temperature Log, page 4-5 •...
  • Page 72: The Run State

    Isco 5800 Refrigerated Sampler Section 4 Operation If the program is set to use a start time, the sampler will wait for the specified time and date. During this delay, the display for time paced programs will show: FIRST SAMPLE...
  • Page 73: Pause Or Stop A Running Program

    Isco 5800 Refrigerated Sampler Section 4 Operation 4.2 Pause or Stop a Press Stop to pause a running program. The sampler will display: Running Program RESUME IN MM:SS VIEW LOG HALT The top line of the display shows a five minute count down timer.
  • Page 74 Isco 5800 Refrigerated Sampler Section 4 Operation • Exit the log by pressing the Stop button. The log generally contains the following information: 1. The log reports the number of successful samples out of the COLLECTED total number programmed. ___ OF ___ SAMPLES 2.
  • Page 75: Retrieve The Temperature Log

    Isco 5800 Refrigerated Sampler Section 4 Operation Table 4-1 5800 Errors Message Description The sampler was unable to detect liquid. NO LIQUID DETECTED After the sampler detected liquid and while the NO MORE LIQUID sample was being taken, the liquid detector stopped detecting liquid.
  • Page 76: Sampler To Rs-232 Serial Communication Cable (Db9-Pin Version Shown)

    · 1 stop bit · No flow control · Auto Detect or VT100 terminal emulation. To sampler To computer Figure 4-1 5800 Sampler to RS-232 serial communication cable (DB9-pin version shown) 5800 Sampler,Model,5800,ID,1158325658, Current Time,38980.62200, Run Time,38980.00000, First Sample Time,38980.55000, 38980.61250,...
  • Page 77: Sampling Reports

    Isco 5800 Refrigerated Sampler Section 4 Operation All times are reported as serial values, which are easily converted to dates and times by spreadsheet programs. The numbers to the left the decimal point are the number of days since January 1, 1900. The numbers to the right are fractional days which can be converted to hours, minutes, and seconds.
  • Page 78: Grab Samples

    Isco 5800 Refrigerated Sampler Section 4 Operation Guide Pin Figure 4-3 Sliding the 24-bottle rack out 4.4 Grab Samples Grab samples let you take a single sample on demand and collecting it in an external container. You can collect a grab sample while the sampler is running a program, paused, or at the Main Menu screen.
  • Page 79: Calibrate Sample Volumes

    (Section 3.3.4). These steps also require a graduated cylinder to measure the delivered sample volume. Note Teledyne Isco offers a 1000 ml graduated cylinder. Order part number 299-0020-00. To calibrate the sample volume: 1. At the Main Menu, press the Calibrate button.
  • Page 80: Operate The Pump Manually

    Isco 5800 Refrigerated Sampler Section 4 Operation For small volumes, repeat the calibration to ensure that the calibration is averaged and not unduly influenced by normal random variations. An alternative method using manually defined sample volumes may be helpful for difficult applications. See Section 3.5.10.
  • Page 81: Periodic Maintenance Checklist

    This section describes how to maintain and troubleshoot your sampler. If you think your sampler requires repair, or if you have questions concerning its operation or maintenance, contact your authorized Isco service facility or Teledyne Isco’s Technical Service Department: Phone: (800) 228-4373 (USA, Canada, Mexico)
  • Page 82: Pump

    Isco 5800 Refrigerated Sampler Section 5 Maintenance WARNING The sampler has a safety interlock that prevents the pump from operating when the pump housing band is open. DO NOT tamper with the pump housing and band. The pump is extremely powerful. The pump can injure you severely if the sampler activates the pump while you are working on it.
  • Page 83: Pump Tube Replacement

    Isco 5800 Refrigerated Sampler Section 5 Maintenance Old Pump Rotor New Pump Rotor Figure 5-2 Rollers (A) and guides (B) on the pump rotor b. Pump Rotor – Look for debris build-up on the pump roller, guide surfaces, and pump band (not shown) (Figure 5-2).
  • Page 84 Isco 5800 Refrigerated Sampler Section 5 Maintenance 2. Loosen the Liquid Detector Cover (F) by unscrewing the large black knob (H). 3. Unlatch the Pump Housing Band (J). 4. Pull the tube (A) away from the bulkhead fitting (B) and disconnect the tubing coupler and suction line.
  • Page 85: Cleaning The Pump Rollers

    Isco 5800 Refrigerated Sampler Section 5 Maintenance Example: Resetting the Pump Tube Alarm 1. From the Main Menu screen select CONFIGURE. 5. You can modify the pump tube alarm setting to tailor it to your needs. The factory default setting is...
  • Page 86: Distributor Calibration

    Isco 5800 Refrigerated Sampler Section 5 Maintenance For application-specific requirements, consult with your laboratory to establish cleaning or replacement protocols. Figure 5-3 Wetted Parts Strainer (316 stainless steel, Bulkhead Fitting (316 stain- polypropylene, or CPVC) less steel) Suction Line (vinyl)
  • Page 87: Sampler Diagnostics

    TEST 'RAM' 2. Wait for the sampler to complete the test..TESTING 'RAM' 3. When finished, the sampler displays the test results. 'RAM' PASSED If the sampler reports ‘RAM’ FAILED contact your authorized Isco service facility or the Teledyne Isco factory.
  • Page 88: Test 'Rom

    'ROM' PASSED If the sampler reports ‘ROM’ FAILED contact your authorized Isco service facility or the Teledyne Isco factory. For these samplers, the test will report NO CHECKSUM. This result does not indicate a pass or fail state, only that the ROM test is not valid for this particular sampler.
  • Page 89: Test Pump

    TEST PUMP diagnostic again. If you don’t find obstructions to pump rotor movement, contact your authorized Isco service facility or the Teledyne Isco factory. 5.2.6 Test Distributor The TEST DISTRIBUTOR diagnostic tests the sampler’s...
  • Page 90: Test Flow Meter Port

    Isco service facility or the Teledyne Isco factory. They will be able to assist you with o b t a i n i n g a t e st p l u g o r m a k i n g t h e n e c e s sa r y j u m p e r connections to run the test.
  • Page 91: Pcb Temperature

    PCB TEMPERATURE: __._ C If you suspect that the internal heater is not working, contact your authorized Isco service facility or the Teledyne Isco factory. 5.2.10 Input Voltage The INPUT VOLTAGE diagnostic displays the Main CPU board’s DC voltage supplied by the power module. When this test is started, the sampler will briefly display the input power voltage.
  • Page 92: Loaded Battery Volts

    Isco 5800 Refrigerated Sampler Section 5 Maintenance The BATTERY VOLTAGE diagnostic displays no-load DC voltage supplied by the internal battery. When this test is started, the sampler will briefly display the battery voltage. To display the no-load battery voltage: 1. Select the BATTERY VOLTAGE option and press Enter.
  • Page 93: Analog Input

    Isco 5800 Refrigerated Sampler Section 5 Maintenance 5.2.13 Analog Input The ANALOG INPUT diagnostic displays the 4-20 mA current loop input to the sampler. This diagnostic tool is helpful when troubleshooting an analog input from a flow measurement device. When you start this test, the sampler will display the current reading until you press the Stop or Enter button.
  • Page 94: Temperature Verification

    Isco 5800 Refrigerated Sampler Section 5 Maintenance 5.3 Temperature Should there be a need to verify the temperature readings displayed by the control panel, it is important to read the Verification temperature as close as possible to the sensor inside the refrigerator.
  • Page 95: Appendix A Replacement Parts List

    Replacement parts are called out in the following pages. Refer to the call-out in the adjacent table to determine the part number for the item. Replacement parts can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco Customer Service Department P.O.
  • Page 96: 5800 Refrigerated Sampler

    Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A.1 5800 Refrigerated Sampler...
  • Page 97 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 98 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 99 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 100 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 101 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 102 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 103 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List...
  • Page 104 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-10...
  • Page 105 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-11...
  • Page 106 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-12...
  • Page 107 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-13...
  • Page 108 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-14...
  • Page 109 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-15...
  • Page 110 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-16...
  • Page 111 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-17...
  • Page 112 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-18...
  • Page 113 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-19...
  • Page 114 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-20...
  • Page 115 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-21...
  • Page 116 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-22...
  • Page 117 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-23...
  • Page 118 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-24...
  • Page 119 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-25...
  • Page 120 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-26...
  • Page 121 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-27...
  • Page 122 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-28...
  • Page 123 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-29...
  • Page 124 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-30...
  • Page 125 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-31...
  • Page 126: Bottle Configurations

    Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A.2 Bottle Configurations A-32...
  • Page 127 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-33...
  • Page 128 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-34...
  • Page 129 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-35...
  • Page 130 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-36...
  • Page 131 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-37...
  • Page 132 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-38...
  • Page 133 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-39...
  • Page 134 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-40...
  • Page 135 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-41...
  • Page 136 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-42...
  • Page 137 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-43...
  • Page 138 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-44...
  • Page 139 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-45...
  • Page 140 Isco 5800 Refrigerated Sampler Appendix A Replacement Parts List A-46...
  • Page 141: Appendix B General Safety Procedures

    Isco 5800 Refrigerated Sampler Appendix B General Safety Procedures The safety of the personnel who use the sampler is a critical consideration. The following procedures, applicable to working in and around manholes and sewers, are those used by Black &...
  • Page 142: Planning

    Isco 5800 Refrigerated Sampler Appendix B General Safety Procedures This applies to hand tools as well as stones, gravel and other objects. • Sharp Edges–Sharp edges of items in or near a manhole may cause cuts or bruises. • Lifting Injuries–Unless proper tools are used to remove manhole covers, back injuries or injuries to hands or feet may result.
  • Page 143: Traffic Protection

    Isco 5800 Refrigerated Sampler Appendix B General Safety Procedures A person should never enter a manhole unless he is wearing personal safety equipment, including a safety harness and a hard hat. Two persons should be stationed at the surface continuously while anyone is working inside a manhole, to lift him out if he is overcome or injured.
  • Page 144: Emergencies

    Isco 5800 Refrigerated Sampler Appendix B General Safety Procedures B.4.4 Emergencies Every member of the crew should be instructed on procedures to be followed in cases of an emergency. It is the duty of each crew chief to have a list of emergency phone numbers, including the nearest hospital and ambulance service, police precinct, fire station, and rescue or general emergency number.
  • Page 145 Isco 5800 Refrigerated Sampler Appendix B General Safety Procedures Whitney, oxygen is equal to 12.2%. Many hikers go there, and to higher peaks as well. After adequate acclimation, they may climb to the top of Mt. Everest, where oxygen is equal to only 6.7%.
  • Page 146 Isco 5800 Refrigerated Sampler Appendix B General Safety Procedures concentrations. A supposed safety device that is unreliable is worse than none at all. “Remember that the nose fails, too, when it comes to sensing dangerous concentrations of H “Various other toxic gases have been mentioned in some publications.
  • Page 147 CE Declaration of Conformity...
  • Page 149 Name and amount of Hazardous Substances or Elements in the product Hazardous Substances or Elements Component Name (Pb) (Hg) (Cd) (Cr(VI)) (PBB) (PBDE) Circuit Boards Display Wiring Internal Cables Compressor Insulation Insulation Foam Line Cord Power Supply DC Motor Keypad Name and amount of Hazardous Substances or Elements in the product O: Represent the concentration of the hazardous substance in this component’s any homogeneous pieces is...
  • Page 151 * This warranty applies to the USA and countries where Teledyne Isco does not have an authorized dealer. Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their Teledyne Isco dealer for warranty service.