Isco 5800 Refrigerated Sampler
Section 3 Programming
Figure 3-5
Suction Head Height
3.5.19 Pause & Resume
3-22
The Pause & Resume option allows you to pause the operation of
the sampler and then resume operation at a later time and/or
date. This option has to be enabled before the sampler is
programmed. It typically runs on continuous mode.
To enable the Pause & Resume option on the sampler:
1. Turn on the sampler.
2. At the Home screen press 5 8 0 0 9.
3. An option screen will appear. Press 1 7 6 0 and Enter.
A screen will appear that reads "Option 1760 Enabled".
Once the Home screen reappears, you can begin programming as
usual. The Pause & Resume programming screen will appear
after the Sample Volume programming is complete.
The first time you get to the Pause & Resume screen it will not
have any information entered, so you will need to select YES to
modify the information by using the arrow button and hitting
Enter. You can enter up to nine Pause & Resume pairs.