Part Number
Description
68-5800-010
1-bottle Configuration. Includes one glass 2.5-gallon (10-liter) round bottle, two PTFE lined caps,
locating base, and two discharge tubes.
68-5800-018
Single ProPak Configuration for 5800 and 6700 Series FR Refrigerated Sampler. Includes one
composite ProPak holder with cap, 100 ProPak bags, two discharge tubes and an instruction sheet.
68-5800-011
1-bottle Configuration. Includes one polyethylene 20-liter round bottle, two caps and two discharge
tubes.
68-5800-012
1-bottle Configuration. Includes one glass 20-liter round bottle, two silicone rubber stoppers and
two discharge tubes.
2.5.1 Installing the
Distributor Arm and
Discharge Tube
Table 2-1 Bottle Kits
If you have selected a multi-bottle kit, first install the distributor
arm and discharge tube (Section 2.5.1), then refer to the
instructions for your selected bottle configuration:
• 24 Bottle configurations – Section 2.5.2
• 4 10L Bottle configurations – Section 2.5.3
• 2 Bottle configurations – Section 2.5.6
• 4 20L Bottle configurations – Section 2.5.5
If you have selected a single (composite) bottle kit, refer to the
instructions in Section 2.5.6.
For first-time use of the sampler, or if you have changed the
bottle kit, be sure to configure the software for the new bottle kit
(Section 3.3.2).
The sampler is shipped with the distributor arm installed on its
mounting shaft inside the refrigerator (Figure 2-2).
1. Thread the discharge tube through the body of the
distributor arm.
2. Push the end of the discharge tube up to the outlet hole on
the distributor arm. The end of the tubing should be flush
with, or within 3 mm (
The discharge tube should not extend more than 3 mm (
below the hole in the distributor arm. This could cause the
distributor to fail if the excess tubing catches a bottle.
3. Pull on the loose end of the discharge tube to remove any
slack. The tube should follow the contour of the distributor
arm.
4. Push the free end of the discharge tube through the tubing
arm and onto the bulkhead fitting in the top-front corner of
the refrigerator.
Isco 5800 Refrigerated Sampler
Section 2 Installation
1
/
") of, the lower edge of the hole.
8
CAUTION
1
/
")
8
2-7