Minimum Space Requirements - Teledyne 5800 Installation And Operation Manual

Refrigerated sampler isco
Hide thumbs Also See for 5800:
Table of Contents

Advertisement

Figure 2-1

Minimum space requirements

thereby reducing the possibility of sample
cross-contamination. Refer to Table 1-2 for maximum
suction line lengths and suction head heights.
• Security – The sample compartment and pump and
control panel have a padlock feature, but the location
may need additional security to prevent tampering or
vandalism.
• Accessibility – The sampler must be installed in a
location where personnel can easily exchange bottles
and perform routine service.
The sampler requires about 2600 square centimeters (925
2
in
) of floor space. Additional space must be allowed in
front of the sampler, at its left and right sides, and above
the sampler. Refer to Figure 2-1. Do not block access to
these areas. Obstructions will make routine servicing
activities difficult or impossible.
The back of the sampler may be placed against a wall as
long as air at the top and bottom of the refrigerator can
flow freely to surrounding air. Unrestricted air flow behind
and around the sampler will allow the refrigeration system
work efficiently.
Sampler dimensions
Keep areas accessible
Door
Latch
85 cm
20 cm
31 cm
61 cm
Door
• Level surface – The sampler should be placed on a level
surface and the feet on the refrigerator body should be
adjusted to prevent tipping or spills. If the sampler is
not level, the sample liquid may miss the bottle mouth.
Isco 5800 Refrigerated Sampler
Section 2 Installation
32 cm
Control
32 cm
Panel
Access
26 cm
133 cm
Pump
Access
73 cm
2-3