Connecting Power; Backup Power - Teledyne 5800 Installation And Operation Manual

Refrigerated sampler isco
Hide thumbs Also See for 5800:
Table of Contents

Advertisement

2.4 Connecting Power

2.4.1 Backup Power

Before connecting the sampler to an AC power source, be
familiar with the Electrical Requirements listed on page v
at the front of this manual.
The factory assembles the sampler for either 100VAC/50 Hz,
115 VAC/60 Hz, or 230 VAC/50 Hz. The serial number label
located inside the refrigerator identifies the AC power
configuration of your sampler. You cannot convert the sampler in
the field without purchasing a replacement refrigeration module.
The mains power cord for 115 VAC samplers is equipped with a
NEMA 5-15P plug for standard North American outlets. The
mains power cord for 230 VAC samplers is equipped with an
EU-16P plug for standard CEE-7 European outlets. For other
types of outlets that supply compatible AC power, convert the
mains power cord plug with a locally-purchased adapter.
Plug the mains power cord into the outlet. The sampler control
panel will briefly show the start up displays.
Copyright 2006 – 2015
COMPRESSOR WARM-UP!
Note
The refrigeration system may not operate for 45 minutes after
applying power. This delay allows time for the compressor
heater to drive off any refrigerant dissolved in the compressor
oil.
The 5800 is capable of switching the sampler controller to backup
power from an Isco lead-acid battery in the event of AC power
loss.
Note
This provides backup power to the sampler controller only; not
to the refrigerator.
To enable power backup, connect the module (60-5314-696) to the
16-pin connector on the back of the sampler (Figure ). Connect
the Isco lead-acid battery to the 2-pin amphenol connector on the
module. The module duplicates the 16-pin connector to allow
connection to other devices.
Isco 5800 Refrigerated Sampler
Section 2 Installation
WARNING
TELEDYNE ISCO
COOLING IN: 44:45
5800 SAMPLER
2-5