Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle (Continued)
SAMPLE VOLUME:
11
200 ml (10-1000)
NO DELAY TO START
DELAYED START
12
CLOCK TIME
WAIT FOR PHONE CALL
FIRST SAMPLE
AFTER A
13
5 MINUTE DELAY
(1-999)
PROGRAMMING COMPLETE
RUN THIS PROGRAM
14
NOW?
YES NO
RUN
PROGRAM
15
VIEW REPORT
OTHER FUNCTIONS
Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample
6712 SAMPLER
STANDARD PROGRAMMING
1
For HELP at any
screen press ? key.
RUN
PROGRAM
2
VIEW REPORT
OTHER FUNCTIONS
SITE DESCRIPTION:
"FACTORY051"
3
CHANGE?
YES NO
Type the volume of the sample you want deposited in each bottle. Then,
press
.
For this example, select DELAYED START by pressing an arrow until the
option blinks. Then, press
Type the delay period you want between the time you run the program and
the time the sampler takes the first sample. Then, press
Run the program immediately by selecting YES. Select NO if you want to
run the program later by selecting RUN from the main menu. Press
making your choice. In this example, NO is selected.
Run the program by selecting RUN and pressing
The following table is an example of flow-paced sampling.
• Program type: Standard
• Site description: SITE 29
• Bottle kit: 24, 1000 milliliter bottles
• Suction line length: 7 feet
• Pacing: Flow pacing, every two pulses
• Distribution: 2 bottles per sample
• Sample Volume: 250 milliliters
• Start time: Clock Time, 6:00 a.m. on Monday,
Wednesday, Friday
• Module: No module installed
Select PROGRAM.
Select NO.
6712SR Refrigerated Sampler
Section 4 Standard Programming
.
.
.
after
4-3