Isco 5800 Refrigerated Sampler
Section 5 Maintenance
MOVING TO BOTTLE ___
NOW AT BOTTLE ___
GO TO BOTTLE ___
(1-max)

5.2.7 Test Flow Meter Port

SELECT DIAG: (<-->)
TEST FLOWMETER PORT
INSTALL TEST PLUG
PRESS
WHEN READY!
FLOWMETER PORT TEST:
...TESTING

5.2.8 Refrig Temperature

5-10
3. The sampler reports the new distributor arm position.
Verify that the arm is aligned over the correct bottle. If the
arm is not aligned over the bottle, the distributor requires
calibration. Contact your authorized Isco service facility or
the Teledyne Isco factory.
4. The sampler then asks for the next bottle position. Enter a
bottle number and press Enter. Or, press the Stop button to
exit the test.
If the sampler displays DISTRIBUTOR JAMMED, DISTRIBUTOR
ERROR, SLOT CODE ERROR, DISK ERROR, or BELT TOO
LOOSE, verify that the arm movement is not obstructed. Clear
any obstructions such as bottles not fully seated in the rack or an
incorrectly installed discharge tube, then perform the TEST
DISTRIBUTOR diagnostic again from step 1. If you don't find
obstructions to the distributor arm movement, contact your
authorized Isco service facility or the Teledyne Isco factory.
The TEST FLOWMETER PORT diagnostic tests the sampler's
ability to send event marks, and receive an enable pin signal and
flow pacing pulses.
Note
This diagnostic routine is not for field use. This test requires a
test plug which is inserted into the External Device connector.
If you have checked all the connections with the external flow
meter and still suspect a fault with the event mark or flow pacing
signals, contact your authorized Isco service facility or the
Teledyne Isco factory. They will be able to assist you with
o b t a i n i n g a t e st p l u g o r m a k i n g t h e n e c e s sa r y j u m p e r
connections to run the test.
To run the test:
1. Select the TEST FLOWMETER PORT option and press
Enter.
2. Insert a Flow Meter Port Test Plug into the Flow Meter
connector.
3. The sampler tests the send and receive circuits for the
event marks, sampler enable pin (inhibit), and flow pulses.
If the sampler displays FAILED: EM->FP or FAILED: INHIBIT,
verify that test plug is correctly inserted. Then, perform the
TEST FLOWMETER PORT diagnostic again from step 1. If the
test still fails, contact your authorized Isco service facility or the
Teledyne Isco factory.
T h e R E F R I G T E M P E R AT U R E d i a g n o s t i c d i s p l a y s t h e
temperature of the refrigerated compartment. When this test is
started, the sampler will display the temperature until you press